SPECIFICATIONS
2-1 General Specifications
Display: 3 3/4 Digit LCD, max. reading of 3999.
Overrange Indication : “OL” indicated.
Measuring Rate: 4/sec, nominal.
Low Battery Indication:
. Replace battery immediately.
Auto Power Off: Approx. 30 minutes after no function change.
2-2
Environmental Conditions
Operating Temperature: 0 °C to +45 °C, <75% R.H.
Storage Temperature: -20 °C to +60 °C, <80% R.H., battery removed.
Temperature Coefficient: 0.2 x (Spec.Acc’y)/°C, <18 °C or >28 °C.
Power Requirement: Single 9V battery (NEDA 1604,1EC 6F22)
Battery Life: Alkaline 200 hours.
Shock Proof: 1.2m (4 feet) drops.
Maximum Jaw Opening: 42mm (1
5
/
8
inches)
Maximum Conductor Size: 40mm (1.5 inches) diameter.
Size (WxLxH): 88x215x36mm (3.5”x8.5”x1.4”)
Weight: 320gms (including battery)
Accessories: Test leads, battery, manual and carrying case.
Safety: Meets EN61010-1 Cat II 600V; EMC: Meets EN50081-1, EN50082-1
EMC: This product complies with requirements of the following
European Community Directives: 89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low Voltage) as amended by
93/68/EEC (CE Marking). However, electrical noise or intense electromagnetic
fields in the vicinity of the equipment may disturb the measurement circuit.
Measuring instruments will also respond to unwanted signals that may be
present within the measurement circuit. Users should exercise care and take
appropriate precautions to avoid misleading results when making measure-
ments in the presence of electronic interference.
2-3 Electrical Specifications
Accuracy is ±(%r nbr digits) at 23 °C ±5 °C, <75% R.H.
AC Voltage (Autoranging)
Range
Resolution
Accuracy (40Hz - 500Hz)
400V
100mV
±(1.2%rdg + 5dgt)
600V
1V
±(1.2%rdg + 5dgt)
AC Conversion Type:
Model AC60: average sensing, rms indication.
Model AC65: AC conversions are ac-coupled, true rms responding. The
basic accuracy is for sine wave at full scale. For Crest Factor of 1.4 to 2.0,
add 0.6% to accuracy; for Crest Factor of 2.0 to 2.5, add 2.0% to accuracy.
Input Impedance: 10M
Ω
with <100pF.
Overload Protection: 600V rms
– 10 –
AC60,AC65.Man.07.00 9/11/00 8:38 PM Page 10