APPENDIX C –
Preprogrammed Measurement Limits
The following Charts show the preprogrammed measurement limits for each
OE-2
model. See Section 3 for more
information on the measurement limit presets.
Table C1. OE-2 Option 1
Protocol
Filter Wavelength
Power Units
Average Power
Limit
Modulation
Measure
Modulation
Units
Modulation
Limit
Rise/Fall times
Limits (in ps)
“User Defined”
Defaults
“Filter 1”
850 nm
dBm
(1)
OMA
uW
(1)
(1)
Gigabit Ethernet
1000BASE-SX
GE 850
nm dBm -9.5 ER dB 9
260 rise
210 fall
Gigabit Ethernet
1000BASE-LX
Single Mode
GE
1300 nm
dBm
-11.0
ER
dB
9
260
Gigabit Ethernet
1000BASE-LX
Multimode
GE 1300
nm
dBm -11.5 ER dB 9 260
Fibre Channel
100-SM-LC-L
FC
1300 nm
dBm
-9.5
OMA
mW
(4)
320
Fibre Channel
200-SM-LC-L
FC2X
1300 nm
dBm
-11.7
OMA
mW
(4)
160
Fibre Channel
100-M5-SN-I
FC 850
nm dBm -10 OMA
mW
0.156 300
Fibre Channel
100-M6-SN-I FC
850
nm
dBm
-10
OMA
mW
0.156
300
Fibre Channel
200-M5-SN-I
FC2X 850
nm dBm
-10
OMA mW 0.196 150
Fibre Channel
200-M6-SN-I FC2X 850
nm
dBm
-10
OMA
mW
0.196
150
Infiniband
1X-SX IB 850
nm
dBm (2) OMA
mW
0.196
150
Infiniband
1X-LX
IB
1300 nm
dBm
(2)
OMA
mW
0.186
150
Infiniband
4X-SX
IB 850
nm dBm (2) OMA
mW
0.150
150
Infiniband
12X-SX
IB 850
nm dBm (2) OMA
mW
0.150
150
(1) Leave blank for user setting
(2) Not specified or multiple specs in the standard. The user must establish the limit.
(3) The Fibre Channel standard offers a tradeoff between spectral center wavelength, RMS spectral width, and
minimum OMA. The user must establish the limit.
APPENDIX C | Preprogrammed Measurement Limits |
27