Interface Requirements
The optical inputs are high return loss FC/UPC connectors. Either single mode or multi-mode fibers can
be accepted.
The electrical outputs are SMA connectors.
The OE-2 includes an IEEE-488 interface. The connector and address selection switch are located on the
rear panel of the instrument.
Power Requirements
The OE-2 operates on 100-120 VAC & 200-240 VAC (±10%), 47-63Hz. The OE-2 is rated for 275 Watts.
Fuses: (2) two 2AT/250V, 5 x 20 mm.
Power supply is auto ranging.
The AC input appliance coupler or separable plug on the power cord is the disconnect device. For safe
operation, the OE-2 should be installed to allow easy operator access to the disconnect device.
Communications Requirements
The OE-2 uses English-language IEEE-488.2 commands. The command list is shown in Appendix III.
Environmental Requirements
Temperature Range
Operating .......................... +15
°
C to +35
°
C
Storage.............................. -40
°
C to + 55
°
C
Humidity.................................. 0-85% RH, non-condensing
Ventilation -
A
minimum of two inches of unobstructed airflow must be maintained on each side and rear
panel of the instrument (air inlet and exhaust openings) to provide adequate ventilation for the
OE-2. To maintain the highest accuracy, the OE-2 air intake should not be near sources that
would cause ambient temperature fluctuations. Temperature fluctuations greater than 5°C will
require that OE-2 be recalibrated.
Elevation (Altitude) ................. to 6500 feet (2000 meters)
Indoor Use Only ...................... The OE-2 is intended for indoor use only.
Installation Category................ CAT II
Pollution Degree ...................... 2 per IEC 664
Product Life Requirements
Product lifetime is five years minimum with recommended factory calibration intervals of one year.
APPENDIX A | Specifications | 21