Limitations and Timeout
The XML commands from the PC to the
SMP2
can be up to 32768 bytes. The
SMP2
expects to receive complete commands in less than 3 seconds (counting
from the time of receipt of the first byte); If after this time it does not have a
full and valid response, an error message is sent and discards the information
received so far. If you send a message without the initial “ < “ XML tag, the
SMP2
will not send any response.
When connecting the USB to the computer, the
SMP2
will interrupt listening the
fibre port.
Example of Commands
These examples have been automatically generated by the software
SMP2
and should be sent unformatted (ie no spaces and line breaks between tags).
System info
<?xml version=”1.0” encoding=”ISO-8859-1”?><cmd><instruction>INFO_SYSTEM</
instruction><md5>c1c0060ec77a5b477c8e164012df8e4f</md5></cmd>
SMP2 time
<?xml version=”1.0” encoding=”ISO-8859-1”?><cmd><instruction>GET_TIME</
instruction><md5>81c73f231ba30eca41b76577ad6f7f9f</md5></cmd>
Field level
<?xml version=”1.0” encoding=”ISO-8859-1”?><cmd><instruction>GET_SAMPLE</
instruction><md5>d8dd6be74a3df920ca20b397a0f49bc0</md5></cmd>
FFT Total
<?xml version=”1.0” encoding=”ISO-8859-1”><cmd><instruction>GET_FFT</
instruction><md5>a03a5727db05e5662a096c1bdf208dc4</md5></cmd>
User’s Manual
Electromagnetic Field Meter SMP2
P.
87
Summary of Contents for SMP2
Page 2: ......
Page 49: ...Figure 11 Installation of the program User s Manual Electromagnetic Field Meter SMP2 P 47...
Page 79: ...Screenshot 4A Screenshot 4B User s Manual Electromagnetic Field Meter SMP2 P 77...
Page 95: ......
Page 96: ...www wavecontrol com...
Page 97: ...Safety Quality Service info wavecontrol com www wavecontrol com 34 93 320 80 55...