Note:
If data is downloaded to a PC, the FFTs for all the axes and the total field can
be displayed in a single graph.
To save an FFT, go to the main screen (direct access available by pressing the
“Home“ button) and press the “LOG“ function. The device will save the current FFT.
6.1.2.1. Case scenarios
This section sets out recommendations for using the selective system in
different situations:
• Checking harmonic content of power lines:
Select:
• “MODE“: FFT
• “SPAN“: 400 Hz
• “FILTER“: 10 Hz
You will obtain an FFT showing 50 Hz and the main harmonics within the
frequency domain (10 to 400 Hz). You can switch between the electric and
magnetic fields using the ‘FIELD’ option.
• Measurement of power lines:
Select:
• “MODE“: Time
• “FILTER“: 10 Hz
In this mode you will obtain the total field value, as with a broadband probe.
Repeat for the magnetic and electric fields.
• Measurement when the field type is unknown:
• We recommend always using a filter of 10 Hz or higher, because
below 10Hz any movement of the device or in its vicinity can affect the
measurement. If you are not sure that there is no field below 10 Hz, use a
filter of 1 Hz, and in this case always take the measurement with the device
set on a tripod to prevent movement.
User’s Manual
Electromagnetic Field Meter SMP2
P.
39
Summary of Contents for SMP2
Page 2: ......
Page 49: ...Figure 11 Installation of the program User s Manual Electromagnetic Field Meter SMP2 P 47...
Page 79: ...Screenshot 4A Screenshot 4B User s Manual Electromagnetic Field Meter SMP2 P 77...
Page 95: ......
Page 96: ...www wavecontrol com...
Page 97: ...Safety Quality Service info wavecontrol com www wavecontrol com 34 93 320 80 55...