Figure 6: Main screen

Total field

Instantaneous total field level. Obtained, for three-axis probes as:

The quadratic sum of each axis field (for field intensity units: V/m, kV/m):

2

2

2

Z

Y

X

Tot

E

E

E

E

+

+

=

The sum of each axis field (for power density units: μW/cm

2

, mW/cm

2

, W/cm

2

):

Z

Y

X

T o t

S

S

S

S

+

+

=

This value is updated once per second.

Field unit

:  

Unit of measurement of the field value shown on the display. This unit

applies to the total field, the field on each axis and the maximum, minimum and average
values.

Field on  each axis

:  

The field measured on each axis for isotropic three-axis probes.

Updated once per second.

Maximum and minimum values

:  

Maximum and minimum values measured for the

field. These values are only updated (once per second) when data are entered in the
memory.

Average value

Average value of the field. This value is only updated when logging is

activated and updated depending on the type of averaging and interval chosen.

Type of average

Type of interval and average selected.

GPS

:  

Positional information provided by the GPS receiver. This information is only

displayed when the GPS is switched on.

11

 -

999.99

mW / cm

2

X

673.98

Max

:   99.78   

Min

 :  24.54

Probe: XXXXXXX

Alarm: 2500

Log

24/04/07  17:15:01           Batt:

Y

45.78

Z

25.54

Avg

:   99.78  

[Arith. – 10 min]

GPS

    Lat:    47º58’02,54” N

           Lon:  102º45’22,78” E   Height: 787.3 m

Total field

Field unit

Field on
each axis

Maximum and 
minimum value

Average 

value

Type of average

Date and 

time

GPS data

Battery 

information

Electric field

Summary of Contents for SMP

Page 1: ...SMP Portable Electromagnetic Field Monitoring System User s Manual V3 9 8 WAVECONTROL C Pallars 65 71 08018 Barcelona SPAIN Tel 34 933 208 055 Fax 34 933 208 056 www wavecontrol com ...

Page 2: ...isation of measurements 15 5 6 2 Calculation of averages 15 5 6 3 IS measurement 16 5 6 4 Maximum and minimum field 16 5 6 5 Alarm 16 5 6 6 GPS 17 6 RECHARGING THE BATTERY 18 7 COMMUNICATION SMP PC 19 7 1 INSTALLING THE HARDWARE 19 7 1 1 USB driver installation 19 7 2 DATA DOWNLOAD SOFTWARE 20 7 2 1 CSV exportation 21 7 2 2 GIS web site exportation 22 7 3 REAL TIME DATA VISUALISATION 23 8 FIRMWARE...

Page 3: ... or repair of the equipment must be performed only by qualified personnel Please contact Wavecontrol Never overtighten when screwing the SMP on the tripod since the thread engagement could break down Environmental information Disposal of your old product Your product is designed and manufactured with high quality materials and components which can be recycled and reused When this crossed out wheel...

Page 4: ...i systems etc Those measurements can be made quickly and easily anywhere because the SMP is portable and simple to handle and use The SMP can sample the field level continuously and store that data automatically in its built in memory at user configured intervals with a capacity of up to 62 000 samples The data stored in the memory can easily be downloaded to a computer via a USB port for later pr...

Page 5: ...note before you sign it 2 Check that all the components of the SMP are there The SMP is delivered in a compact briefcase that should contain the following components 1 SMP 2 Electromagnetic field probe 3 Battery charger 4 CD or USB key containing the User s Manual and data download software 5 USB mini USB cable If any of those components is missing contact Wavecontrol 5 SMP and USB cable below Bat...

Page 6: ...low to change from electric to magnetic field and range high or low or vice versa Enter Accepts options on the Menu Memory and Memorised Data screens Also block or unblock field value visualization in main screen see 5 2 Back Exits from the Menu Memory and Memorised Data screens In the main screen reset Max Min and Avg fields to 0 Memory Access to the Memory screen which shows the memory data regi...

Page 7: ...e equipment on and off The bottom of the SMP holds the other connections of the equipment Figure 3 Bottom Mini USB connector Allows connection to a PC for 2 functions o Download data stored in the SMP memory o Update firmware for the equipment Battery charger connector Allows the SMP battery to be recharged using the charger supplied with the equipment 7 On Off switch Probe connector GPS antenna R...

Page 8: ...nected to the device when it is switched on it will not provide any field values and therefore no measurements can be stored in the memory However the configuration memorised data display and GPS positioning functions will be available The device must be switched off when a field probe is connected to it The SMP will automatically detect the type of probe and its calibration 4 2 1 Connecting the p...

Page 9: ...he field probe Pull the connector of the probe firmly with one hand pulling with the thumb while you hold the SMP with your other hand as shown in the pictures below Figure 5 Disconnecting the probe Important Pull the connector Don t pull any other part of the probe Don t try to turn the probe 9 ...

Page 10: ... for a few seconds automatically displaying the main screen afterwards The device s configuration alarm level units type of measurement measurement interval etc is saved when it is switched off and recovered automatically when it is switched on 5 2 Main screen The main screen provides instantaneous and continuous information on the electromagnetic field being measured and also on the most importan...

Page 11: ...um values Maximum and minimum values measured for the field These values are only updated once per second when data are entered in the memory Average value Average value of the field This value is only updated when logging is activated and updated depending on the type of averaging and interval chosen Type of average Type of interval and average selected GPS Positional information provided by the ...

Page 12: ...rent options for each parameter are shown on the right The keypad is used to run through menu The process for changing the value of a parameter is Run through the different parameters using the arrows To change a parameter press Enter when the cursor is on that parameter The cursor will move to the right of the screen where the arrows are used to select the desired option Press Enter again to conf...

Page 13: ...mbol o By pressing the key you gain access to the validation option and the text Sure appears o To validate the erasure press Enter o To exit the option without erasing press Back Reset reg ID Used to reset to 0 the unique identifier for each register Language Used to change the language for the system GPS This enables or disables the GPS receiver To do so select the Enabled or Disabled option the...

Page 14: ...Enter o To exit the option without erasing press Back Use the Back key to exit the memory screen 5 5 Memorised data screen This screen shows a list of the data stored in a given memory register Each memory point contains an electromagnetic field measurement with the following data Date Time Date and time when the field value was measured Field Instantaneous field value showing the field unit Avg A...

Page 15: ... When Enter is pressed again in response to the prompt Ready to measure memorisation of the new register starts Entry of a name for each register created is not strictly necessary since each register has a unique ID number However naming registers is advisable since it facilitates identification of the measurements stored in each register To stop memorisation press the Log RD key again If the memo...

Page 16: ...t a given place without the presence of the technician affecting the result To make this type of measurement select Average type IS on the configuration screen Average interval 6m and Log interval 1s will be selected automatically Under those conditions when the Log RD is pressed 1 The apparatus will wait 10s before starting the measurement to allow the technician to move away After the 10s a beep...

Page 17: ...ation menu as explained in section 5 3 When it is enabled the main screen displays the message Searching for satellites along with a number indicating the number of satellites found When the position obtained by the receiver is valid the data for latitude longitude and altitude are displayed on the screen The best position for the device for finding satellites is vertical with the GPS antenna poin...

Page 18: ...again The battery charge time is 4 5 hours The procedure for charging the battery is 1 Insert the connector of the battery charger into the connector of the SMP 2 Plug the charger into an electrical outlet 220 V AC The recharge LED will light up If it does not light up it means that the charger is not connected properly 3 The level of charge of the battery can be checked directly in the Battery fi...

Page 19: ... the new device 2 Do not allow Windows to search for the driver automatically 3 Select Install from a list or specific location 4 Select D Drivers as the location where D is the name of the CD DVD unit 5 Wait until the driver has been correctly installed If advice that is not a certified driver click on Continue 6 If the driver is correctly installed when the SMP is connected to the PC it will app...

Page 20: ...s Windows XP the Microsoft Net Framework must be installed on the PC If SMP Reader does not work after installation install the executable file that appears in the NET Framework2 NET32 folder of the CD if your PC is 32 bits or the executable file from NET Framework2 NET64 if the PC is 64 bits SMP Reader is a very simple program Its purpose is to connect the PC to the SMP to download the data store...

Page 21: ...t it 7 2 1 CSV exportation You can select the registers to be downloaded by means of the selection box to the left of each register or you can select all the register clicking on sign above all the selection boxes Pressing the Download Selected Register button starts the download with the name of the file indicated by the user A file is created for each register of measurements and it is assigned ...

Page 22: ...u have a valid URL user and password in the Option menu see fig 9 Figure 9 SMP reader main option menu you have selected a IS register with the 360 values In this case click on Send to control center It will show you a new window fig 9 where you can adjust the GPS position or enter an address at the bottom to find the location of your measurement You can fine tune the position by dragging the mark...

Page 23: ...d When OK appears you can close this window This measurement will appear automatically in the web site 7 3 Real time data visualisation This function allows instant visualization of data being measured via the USB port When it is switched on the SMP will automatically send the following data to the PC Time and date Field values Field on the three axes and total field value GPS data Longitude latit...

Page 24: ...To activate this function you will have to send this command to the unit USB_ON with CR LF at the end To deactivate this mode the command is USB_OFF 24 ...

Page 25: ... the apparatus in the Firmware Update With the device switched on press the update button on the back of the SMP reached through a small aperture in the case You will need to use a thin rigid tool to press the reset button In this mode the screen is erased and all the functions of the apparatus are blocked until it is restarted Once the SMP in the Firmware Update mode follow this process 1 Connect...

Page 26: ...umber of the assigned COM port is higher than 8 change it to a port number between 2 and 8 2 Start Fujitsu Flash MCU Programmer 3 Set the COM port by clicking on Set Environment and selecting the port number 26 ...

Page 27: ... D E B P button 7 If the program displays the following message Please reset the microcontroller on userboard Then push OK button just click on Accept 8 When the update is complete the program will display the following message Click on OK 9 To end switch off the SMP disconnect the USB cable and switch the apparatus on again The new firmware has been installed To verify installation check the firm...

Page 28: ...ipes to clean the apparatus and the screen Use extra care when cleaning the screen to avoid scratching it 9 2 Handling the equipment The SMP and its field probes are not watertight When using them outdoors make sure that no moisture enters the apparatus Be sure to protect the apparatus from rain The apparatus and the probes should also be protected from any strong blows or vibrations 9 3 Regular r...

Page 29: ...adownload MiniUSBconnector Firmwareupdate MiniUSBconnector Batterycharger coaxialjackconnector LED Display Displaytype Transmissive monochrome STN negative Displaysize 60x78mm 240x320pixels Backlight WhiteLED Displayrange 5digits 0 0001to99999 GPS General SiRFstarIII 12 channels embedded Chip set SiRF starIII GSC3 Position accuracy 1 5 m CEP50 1 8 m CEP95 Sensitivity 157 dBm tracking SBAS Satellit...

Page 30: ...stributors of the defect before the expiration of the warranty period by sending a Failure Report including Defect Report Equipment data model serial number purchase date Contact data Company name and complete contact details This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Wavecontrol shall not be obligated to furn...

Page 31: ...ts and methods of measurement UNE EN 61000 3 2 2006 A1 2010 A2 2010 Electromagnetic compatibility EMC Part 3 2 Limits Limits for harmonic current emissions equipment input current 16 A per phase UNE EN 61000 3 3 2009 Limitation of voltage changes voltage fluctuations and flicker in public low voltage supply systems for equipment with rated current 16 A per phase and not subject to conditional conn...

Reviews: