background image

6

3.0 Theory of Operation

3.1 The Measurement

The CLX-Ex has two solenoid valves, one for sample water (FLOW) 

and one for draining of the cuvette (PURGE). A third solenoid, along 

with four check valves forms a reagent pump. Sample water flow is 

controlled by the FLOW solenoid valve. The PURGE solenoid valve is 

used to empty the cuvette in the measurement chamber.
The measurement chamber consists of a sample inlet, a purge drain, 

and an overflow. The reagent is added from the check valves integrated 

into the lower portion. A green LED provides the 515 nm source 

lamp, a red LED is used for sample level and flow measurement. A 

single detector is located 180º from the green LED. A replaceable 

glass cuvette separates the LEDs from the detector and maintains 

the measurement path length. Sample water flows in the inlet at the 

bottom, through the measurement cuvette and out through an overflow 

drain. This flow is used to both fill the cuvette and flush the system.
The reagents are dispensed from two replaceable bottles. One bottle 

has a buffer to control the pH; the second has an indicator that 

contains the DPD, which produces color when chlorine is present in the 

sample. The degree of color is dependent on the amount of chlorine in 

the sample water.
The measurement chamber is open to view operations. A white LED 

backlights the chamber for a clearer view. The white LED will flash to 

attract attention in the case of a warning or failure. Most warnings and 

failures are also displayed on the screen. To prevent interference, the 

white LED is turned off during measurements.
During normal operation the CLX-Ex will run through a timed cycle. A 

simplified cycle will consist of the following sequences:

•  Flushing – continuous sample flow
•  Purging – PURGE valve opens
•  Zeroing – no flow with cuvette full
•  Adding Reagents – one pulse of the reagent pump
•  Mixing with sample – sample flow pulses in
•  Reading resulting sample – no flow with cuvette full
•  Purging – PURGE valve opens to remove reacted sample

The cycle above is simplified and does not describe all the actions 

and testing that occurs. The CPU continuously diagnoses the entire 

system for correct operation and sample water flow. If an error occurs, 

a message is posted to the message queue on the LCD screen.
The reagent is added by a single pulse of the reagent solenoid. When 

the reagents require replacement, the 

PRIME

 button is pushed to bring 

new reagents into the system. During 

PRIME

 the reagent solenoid is 

pulsed several times to draw fluid from the two reagent bottles and 

fill the tubes with new reagent. A complete 

PRIME

 takes less than a 

minute.
The 

SERVICE

 button empties the cuvette, stops the flow of sample 

water, and clears any errors. This provides a convenient way to replace 

reagents and the measurement cuvette. If more extensive servicing is 

performed, all power to the CLX-Ex should be removed. If the CLX-Ex 

is to be turned off, it is recommended that the instrument be placed in 

SERVICE

 mode before removing power. This ensures that the cuvette 

is emptied and the flow is off.

3.2 The Air Sentinel Controller

The Air Sentinel controller governs the power to the instrument. For 

normal instrument operation air needs to passes through fitting on 

the base of the main enclosure and constantly purge any vapors that 

could build-up. All GREEN lamps indicate a safe operating condition. 

Only when this condition is reached will power be applied to the 

measurement portion of the instrument.

Start-up

At start-up with the enclosure door open, the Purge Control will display 

a RED 

PRESSURE

 lamp on the visual indicator indicates that the 

system is not pressurized and no power is sent to the instrument.
Closing the door with the air pressure applied, will cause the Air 

Sentinel to display a GREEN 

PRESSURE

 lamp and RED 

TIMER

 lamp. 

During this time the enclosure will undergo an automatic dilution purge 

period of at least 5 minutes where at least 5 volumes of free space 

inside the enclosure will be exchanged with clean purge air. No power 

is applied to the rest of the instrument during this time.
After the enclosure air has gone through this purge period the Air 

Sentinel will display all GREEN lamps. At this time the instrument is 

deemed safe and power is applied to rest of the instrument.

WARNING: Do not apply power to the Air Sentinel controller 

unless the area has been properly tested and is known not to 

contain explosive materials.

Bypass Key

WARNING: Use of the bypass key disables the safety features 

and safe operation of this instrument. It is imperative that this 

operation be performed only after the area has been tested and 

known to be safe.
If power is retained to the CLX-Ex after opening the enclosure 

door while NOT in bypass mode, immediately close the 

enclosure door and remove power to the instrument. Contact 

HF scientific service department for further instruction.

Assuming the area has been tested and is known not to contain 

explosive materials; a bypass service key is provided and can be 

used. In this operating mode power will be retained to the rest of the 

instrument even though the door is open. It is imperative that this mode 

is only used after area is known to be safe. This operation disables the 

safe operation of this instrument.
Upon turning the key, the indicator will display the RED 

BYPASS & 

TIMER

 lamps. This is an indication of the maintenance mode. Power is 

retained to the instrument. The key cannot be removed in this mode of 

operation.
After maintenance has been performed the door should be closed 

and latched. Air pressure will once again pressurize the enclosure. The 

maintenance key can now be turned and removed and the key cover 

replaced. The purge controller will display the GREEN 

PRESSURE & 

BYPASS

 lamps indicating a safe operation and power is kept applied to 

the instrument. The 

TIMER

 lamp will be GREEN immediately. No purge 

time is required since door was opened under a non-hazardous condition.

3.3 Remote Standby

This feature will allow for remotely placing the instrument into a standby 

mode. This can be done to lower reagent usage while not performing 

ballasting or de-ballasting operations. The last operation of each 

cycle is a flush and hold. Whenever the instrument is not being used 

(standby), it is stored such that the optics is clean and dry.

Dry contact are provided as feedback as to the operation of the 

instrument (operating or standby). For more information on this 

feature refer to sections 4.4.4 

Remote Standby Connections 

and 6.2 Routine Measurement with Remote Standby.

WARNING

!

Summary of Contents for scientific CLX-Ex A28031

Page 1: ...ended Reagent Life Model CLX Ex A28031 IOM HF CLX Ex 29146 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference IOM HF CLX Ex 29146 Rev 4 ...

Page 2: ... Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring the Error Level 13 8 3 Configuring the RS 485 Port 13 8 4 Configuring the Alarm 14 8 4 1 Alarm1 14 8 5 Enabling the Security Access 14 8 6 Extended Settings 14 8 7 Units of Measurement 14 8 8 Averaging and Filtering 14 8 9 LCD...

Page 3: ...sphere where the instrument is located is safe from gases vapors dust or other flammable conditions whenever the instrument is opened for service or maintenance At all other times the door must be kept closed with purge air connected the purge system operating with the all GREEN lamps lit indicating safe operation WARNING If power is retained to the CLX Ex after opening the enclosure door while NO...

Page 4: ...re Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C 32 F 131 F Wetted Materials PVC Borosilicate Glass Reslyn FFKM Viton FKM Polypropylene 316 Stainless Steel Acetal Noryl Silicone Sample Temperature Range 0 C 55 C 32 F 131 F Power Supply 100 to 240VAC 47 63 Hz 250VA Insulation Rating Double Insulated Pollution Degree...

Page 5: ...on enters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to prime the tubing after a change or addition of reagent bottles The SERVICE button will drain the instrument and hold all operations until either the SERVICE button is pushed again or the power is r...

Page 6: ...ent operation air needs to passes through fitting on the base of the main enclosure and constantly purge any vapors that could build up All GREEN lamps indicate a safe operating condition Only when this condition is reached will power be applied to the measurement portion of the instrument Start up At start up with the enclosure door open the Purge Control will display a RED PRESSURE lamp on the v...

Page 7: ... by the Purge Air Sentinel controller to the cooling chamber temperature controller Clean dry compressed air must also be provided at pressures between 5 5 and 7 bar 80 to 101 5 PSI The cooling chamber temperature controller governs the solenoid which controls air flow to the vortex cooling unit The air flow is turned on or off as required to maintain the indicating reagent temperature at 20 25 C ...

Page 8: ...allation of electrical connections to and between the instrument and other peripheral devices All terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 485 The RS 485 half duplex 2 wire digital interface operates with differential levels that are not suscep...

Page 9: ...tion of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 Start Up When power is applied to the CLX Ex for the first time ensure that the air supply is connected as stated in section 4 3 and the cabinet is closed and locked Once power is applied the Air Sent...

Page 10: ... of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM Readings above 10 00 mg L are outside the range of this instrument Although the CLX Ex may display above 10 0 mg L these readings will not be within the stated accuracy As the reagents degrade due to aging readings above 10 0 mg L ma...

Page 11: ... changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the code Proceed as with the first number followed by Then repeat the process for the third number in the access code and finish with the button If the valid access code has been selected the instrument will be directed to t...

Page 12: ...he size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the display before calibrating to avoid a nOnE error 7 2 Zero offset Calibration Procedure Generally this calibration is only required if readings are expected to be below 1 mg L or if it is required by a re...

Page 13: ...The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press the and button to accept it NOTICE To exit the CONFIG mode press the MODE EXIT button any changes that were made will be saved NOTICE The 4MA can be set higher than 20MA level to invert the output current if req...

Page 14: ...iter or PPM parts per million The factory setting is mg L Select the desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex can display and output averaged readings to help smooth out the response and eliminate large reading variation in rapidly changing processes There are 5 settings for the averaging feature 1 No averaging each reading is in r...

Page 15: ...ss the button Allowable setting is from 60 to 600 seconds 10 minutes NOTICE This feature is generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the instrument sits idle for about 40 seconds This results in about a 25 water savings The actual amount of wat...

Page 16: ... modes Included are a simple communication mode and Modbus communications Both modes will automatically configure and do not require any changes or selections 9 1 1 Simple Communication The CLX Ex can provide basic communications over simple programs such as the Hilgraeve HyperTerminal that is included with most Microsoft Windows packages The user could also use Visual Basic or other programs The ...

Page 17: ...e unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queued display will show FAIL the upper display will show a four or five digit fault code If this fault occurs it will affect the 4 20mA and any alarm set for fault detection Error The instrument will not operate with this faul...

Page 18: ...er Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex one for the buffer and one for the indicator Additional kits can be ordered from your local HF scientific distributor or representative It is recommended to keep one kit on hand at all times Generally both pump tubes should be replaced annua...

Page 19: ...the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T strainer is integral to the instrument and must be checked occasionally When necessary it must be removed and cleaned The strainer screen may require replacement after a period of time Steps 1 Press SERVIC...

Page 20: ...nds 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to use only chlorine free water Complete instructions are included in the kit Figure 6 Check Valve Flushing 11 5 Instrument Storage If the CLX Ex is relocated or will be inactive for long periods of tim...

Page 21: ...ies and replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Purge Manual CLX Ex Air Sentinel 28749 N A Operating Manual CLX Ex 29146 N A Tubing Cuvette Kit 09950EX Replacement Cuvette 25018S Check Valve Kit 25017S Ferrite for 4 20mA or RS 485 24560 ...

Page 22: ...ies and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrument please do not hesitate to contact the HF scientific Technical Services Department HF scientific 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 Phone 239 337 2116 Toll Free 888 203 7248 Fa...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Page 24: ...for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water con...

Reviews: