17
Please note that these operating screens cannot
be seen except with the cabinet door open and the
instrument in maintenance bypass mode. See section
5.3 for more information on this.
The CLX-Ex performs continuous diagnostic monitoring. In the CLX-Ex,
there are 4 severity levels of fault detection. Level 4, 3 and 2 will allow
normal operation, but warn of the problem. Level 1 is an instrument
failure and the instrument will not operate. Any faults are displayed in a
queue form in the bottom row of the LCD.
A
level 4 fault
is simply a screen indication that one of the alarm levels
has been activated. This fault level will not affect the 4-20 mA and will
only affect the alarm activated. The sample back light blinks at a rate of
once every 4 seconds.
A
level 3 fault
indicates a failure or a problem that usually can be
corrected by the operator. Refer to the chart below. If any of these
errors occur, the instrument will still display readings and probably
will operate correctly. These faults will self-clear when the problem is
corrected. If any of these faults occur, they may affect the 4-20mA and
any alarm dependent on the detection setting (Error). See sections 8.2
and 8.4.1 for error settings. The sample back light blinks at a rate of
once every 2.5 seconds to indicate these faults.
Level 3 (Self-Clearing) fault conditions
Message
Description of Fault
Corrective Action
MA
4-20 mA enabled and loop open
Check 4-20 mA wiring or turn off 4-20mA
if not used
CAL
Calibration invalid – not accepted
Recalibrate if needed
WATR
No water flowing
Check water flow
FAST
Intake water flow too fast
Set flow rate (see section 10.2)
SLOW
Sample cuvette filling too slowly
Set flow rate (see section 10.2)
PURG
Sample cuvette has slow purge
Check drain lines
NPRG
Sample cuvette not purging
Check drain lines
ISOL
Problem with intake solenoid
Check wiring, check for clogged solenoid
PSOL
Problem with purge solenoid
Check wiring, check for clogged solenoid
GLAS
Dirty cuvette
Replace or clean cuvette
WCAL
Water Level Calibration Invalid
Clear fault (see section 10.3)
A
level 2 fault
indicates a severe problem that will usually
require technical assistance from HF scientific customer
service (see section 10.6). The queued display will show
POST
. If this fault occurs it will affect the 4-20mA and any
alarm set for fault detection (Error). The sample back light blinks at rate
of once every 1 second.
A
level 1 fault
is a system fault. This is NOT a problem that
the operator can correct, and the unit must be returned
to the factory for service (see section 10.6). These failures
consist of failures in the CPU, A/D, EEPROM or other
devices internal to the instrument. The queued display will
show
FAIL
, the upper display will show a four or five digit fault code.
If this fault occurs, it will affect the 4-20mA and any alarm set for fault
detection (Error). The instrument will not operate with this fault. The
sample back light blinks at rate of once every 0.4 seconds.
If any fault condition occurs, the message indicating the fault will be
shown on the lower row of the display.
10.2 Setting Flow Rate
The flow rate on the CLX-Ex was factory adjusted and usually should
not need adjustment. Installation variances may affect the flow. The
optimal flow rate through the CLX-Ex may be adjusted if needed.
The flow is adjusted by removing regulator vinyl cap and turning the
adjustment screw on the pressure regulator. To assist in this adjustment
follow the procedure shown below:
1. Press the
SERVICE
button.
2. Wait for the display to read HOLd, then press
Mode/ Exit
.
3. Display will show FLOW with the number 0. Press either the
or
button.
4. CLX-Ex will drain, and then pulse in water while a count is displayed
on the screen.
5. The display will show one of three messages HI, LO or Good.
The flow test determines if the flow rate is suitable for proper operation.
Loosen the locking nut then adjust the pressure regulator using a coin
or a large flat blade screwdriver. Press either the
or
button while
in the FLOW routine to display a new flow rate. Please note that only
¼ turn incremental adjustments should be made to the regulator on
each attempt.
If the message is LO, turn the regulator control clockwise. If the
message is HI, turn the regulator counterclockwise. The optimal setting
is when the number of counts is adjusted to 8-10. Once completed,
tighten the locking nut after adjustment and replace the regulator vinyl
cap. To return to normal operation, press the
button.
10.3 Clearing Faults
Every time
SERVICE
mode is exited, all faults are cleared. If the original
fault or a new fault occurs, it will be posted.
In the case of WCAL, it is possible the power was enabled before a
water sample was available. Ensure a water sample is available and
then press PRIME to force another Water Calibration (WCAL).
10.4 Reagent Clogs
If reagents fail to flow or Prime it may be due to a clog in
either the tubing or at a check valve. To alleviate this you
may have to flush the system with Chlorine Free water,
preferably Deionized water. See section11.4
Check Valve
Flushing Kit.
10.0 Troubleshooting
10.1 CLX-Ex Fault Detection
Only access when the atmosphere has been tested and known to
be safe.
WARNING
!