background image

17

Please note that these operating screens cannot 

be seen except with the cabinet door open and the 

instrument in maintenance bypass mode. See section 

5.3 for more information on this.

The CLX-Ex performs continuous diagnostic monitoring. In the CLX-Ex, 

there are 4 severity levels of fault detection. Level 4, 3 and 2 will allow 

normal operation, but warn of the problem. Level 1 is an instrument 

failure and the instrument will not operate. Any faults are displayed in a 

queue form in the bottom row of the LCD.

level 4 fault

 is simply a screen indication that one of the alarm levels 

has been activated. This fault level will not affect the 4-20 mA and will 

only affect the alarm activated. The sample back light blinks at a rate of 

once every 4 seconds.

level 3 fault

 indicates a failure or a problem that usually can be 

corrected by the operator. Refer to the chart below. If any of these 

errors occur, the instrument will still display readings and probably 

will operate correctly. These faults will self-clear when the problem is 

corrected. If any of these faults occur, they may affect the 4-20mA and 

any alarm dependent on the detection setting (Error). See sections 8.2 

and 8.4.1 for error settings. The sample back light blinks at a rate of 

once every 2.5 seconds to indicate these faults.

Level 3 (Self-Clearing) fault conditions

Message

Description of Fault

Corrective Action

MA

4-20 mA enabled and loop open

Check 4-20 mA wiring or turn off 4-20mA 

if not used

CAL

Calibration invalid – not accepted

Recalibrate if needed

WATR

No water flowing

Check water flow

FAST

Intake water flow too fast

Set flow rate (see section 10.2)

SLOW

Sample cuvette filling too slowly

Set flow rate (see section 10.2)

PURG

Sample cuvette has slow purge

Check drain lines

NPRG

Sample cuvette not purging

Check drain lines

ISOL

Problem with intake solenoid

Check wiring, check for clogged solenoid

PSOL

Problem with purge solenoid

Check wiring, check for clogged solenoid

GLAS

Dirty cuvette 

Replace or clean cuvette

WCAL

Water Level Calibration Invalid

Clear fault (see section 10.3)

level 2 fault

 indicates a severe problem that will usually 

require technical assistance from HF scientific customer 

service (see section 10.6). The queued display will show 

POST

. If this fault occurs it will affect the 4-20mA and any 

alarm set for fault detection (Error). The sample back light blinks at rate 

of once every 1 second.

level 1 fault

 is a system fault. This is NOT a problem that 

the operator can correct, and the unit must be returned 

to the factory for service (see section 10.6). These failures 

consist of failures in the CPU, A/D, EEPROM or other 

devices internal to the instrument. The queued display will 

show 

FAIL

, the upper display will show a four or five digit fault code. 

If this fault occurs, it will affect the 4-20mA and any alarm set for fault 

detection (Error). The instrument will not operate with this fault. The 

sample back light blinks at rate of once every 0.4 seconds.
If any fault condition occurs, the message indicating the fault will be 

shown on the lower row of the display.

10.2 Setting Flow Rate

The flow rate on the CLX-Ex was factory adjusted and usually should 

not need adjustment. Installation variances may affect the flow. The 

optimal flow rate through the CLX-Ex may be adjusted if needed. 

The flow is adjusted by removing regulator vinyl cap and turning the 

adjustment screw on the pressure regulator. To assist in this adjustment 

follow the procedure shown below:
1.  Press the 

SERVICE

 button.

2.  Wait for the display to read HOLd, then press 

Mode/ Exit

.

3.  Display will show FLOW with the number 0. Press either the 

 or 

 

button.

4.  CLX-Ex will drain, and then pulse in water while a count is displayed 

on the screen.

5.  The display will show one of three messages HI, LO or Good.
The flow test determines if the flow rate is suitable for proper operation. 

Loosen the locking nut then adjust the pressure regulator using a coin 

or a large flat blade screwdriver. Press either the 

 or 

 button while 

in the FLOW routine to display a new flow rate. Please note that only 

¼ turn incremental adjustments should be made to the regulator on 

each attempt.
If the message is LO, turn the regulator control clockwise. If the 

message is HI, turn the regulator counterclockwise. The optimal setting 

is when the number of counts is adjusted to 8-10. Once completed, 

tighten the locking nut after adjustment and replace the regulator vinyl 

cap. To return to normal operation, press the 

  button.

10.3 Clearing Faults

Every time 

SERVICE

 mode is exited, all faults are cleared. If the original 

fault or a new fault occurs, it will be posted.
In the case of WCAL, it is possible the power was enabled before a 

water sample was available. Ensure a water sample is available and 

then press PRIME to force another Water Calibration (WCAL).

10.4 Reagent Clogs

If reagents fail to flow or Prime it may be due to a clog in 

either the tubing or at a check valve. To alleviate this you 

may have to flush the system with Chlorine Free water, 

preferably Deionized water. See section11.4 

Check Valve 

Flushing Kit.

10.0 Troubleshooting

10.1 CLX-Ex Fault Detection

Only access when the atmosphere has been tested and known to 

be safe. 

WARNING

!

Summary of Contents for scientific CLX-Ex A28031

Page 1: ...ended Reagent Life Model CLX Ex A28031 IOM HF CLX Ex 29146 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference IOM HF CLX Ex 29146 Rev 4 ...

Page 2: ... Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring the Error Level 13 8 3 Configuring the RS 485 Port 13 8 4 Configuring the Alarm 14 8 4 1 Alarm1 14 8 5 Enabling the Security Access 14 8 6 Extended Settings 14 8 7 Units of Measurement 14 8 8 Averaging and Filtering 14 8 9 LCD...

Page 3: ...sphere where the instrument is located is safe from gases vapors dust or other flammable conditions whenever the instrument is opened for service or maintenance At all other times the door must be kept closed with purge air connected the purge system operating with the all GREEN lamps lit indicating safe operation WARNING If power is retained to the CLX Ex after opening the enclosure door while NO...

Page 4: ...re Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C 32 F 131 F Wetted Materials PVC Borosilicate Glass Reslyn FFKM Viton FKM Polypropylene 316 Stainless Steel Acetal Noryl Silicone Sample Temperature Range 0 C 55 C 32 F 131 F Power Supply 100 to 240VAC 47 63 Hz 250VA Insulation Rating Double Insulated Pollution Degree...

Page 5: ...on enters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to prime the tubing after a change or addition of reagent bottles The SERVICE button will drain the instrument and hold all operations until either the SERVICE button is pushed again or the power is r...

Page 6: ...ent operation air needs to passes through fitting on the base of the main enclosure and constantly purge any vapors that could build up All GREEN lamps indicate a safe operating condition Only when this condition is reached will power be applied to the measurement portion of the instrument Start up At start up with the enclosure door open the Purge Control will display a RED PRESSURE lamp on the v...

Page 7: ... by the Purge Air Sentinel controller to the cooling chamber temperature controller Clean dry compressed air must also be provided at pressures between 5 5 and 7 bar 80 to 101 5 PSI The cooling chamber temperature controller governs the solenoid which controls air flow to the vortex cooling unit The air flow is turned on or off as required to maintain the indicating reagent temperature at 20 25 C ...

Page 8: ...allation of electrical connections to and between the instrument and other peripheral devices All terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 485 The RS 485 half duplex 2 wire digital interface operates with differential levels that are not suscep...

Page 9: ...tion of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 Start Up When power is applied to the CLX Ex for the first time ensure that the air supply is connected as stated in section 4 3 and the cabinet is closed and locked Once power is applied the Air Sent...

Page 10: ... of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM Readings above 10 00 mg L are outside the range of this instrument Although the CLX Ex may display above 10 0 mg L these readings will not be within the stated accuracy As the reagents degrade due to aging readings above 10 0 mg L ma...

Page 11: ... changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the code Proceed as with the first number followed by Then repeat the process for the third number in the access code and finish with the button If the valid access code has been selected the instrument will be directed to t...

Page 12: ...he size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the display before calibrating to avoid a nOnE error 7 2 Zero offset Calibration Procedure Generally this calibration is only required if readings are expected to be below 1 mg L or if it is required by a re...

Page 13: ...The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press the and button to accept it NOTICE To exit the CONFIG mode press the MODE EXIT button any changes that were made will be saved NOTICE The 4MA can be set higher than 20MA level to invert the output current if req...

Page 14: ...iter or PPM parts per million The factory setting is mg L Select the desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex can display and output averaged readings to help smooth out the response and eliminate large reading variation in rapidly changing processes There are 5 settings for the averaging feature 1 No averaging each reading is in r...

Page 15: ...ss the button Allowable setting is from 60 to 600 seconds 10 minutes NOTICE This feature is generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the instrument sits idle for about 40 seconds This results in about a 25 water savings The actual amount of wat...

Page 16: ... modes Included are a simple communication mode and Modbus communications Both modes will automatically configure and do not require any changes or selections 9 1 1 Simple Communication The CLX Ex can provide basic communications over simple programs such as the Hilgraeve HyperTerminal that is included with most Microsoft Windows packages The user could also use Visual Basic or other programs The ...

Page 17: ...e unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queued display will show FAIL the upper display will show a four or five digit fault code If this fault occurs it will affect the 4 20mA and any alarm set for fault detection Error The instrument will not operate with this faul...

Page 18: ...er Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex one for the buffer and one for the indicator Additional kits can be ordered from your local HF scientific distributor or representative It is recommended to keep one kit on hand at all times Generally both pump tubes should be replaced annua...

Page 19: ...the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T strainer is integral to the instrument and must be checked occasionally When necessary it must be removed and cleaned The strainer screen may require replacement after a period of time Steps 1 Press SERVIC...

Page 20: ...nds 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to use only chlorine free water Complete instructions are included in the kit Figure 6 Check Valve Flushing 11 5 Instrument Storage If the CLX Ex is relocated or will be inactive for long periods of tim...

Page 21: ...ies and replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Purge Manual CLX Ex Air Sentinel 28749 N A Operating Manual CLX Ex 29146 N A Tubing Cuvette Kit 09950EX Replacement Cuvette 25018S Check Valve Kit 25017S Ferrite for 4 20mA or RS 485 24560 ...

Page 22: ...ies and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrument please do not hesitate to contact the HF scientific Technical Services Department HF scientific 16260 Airport Park Drive Suite 140 Fort Myers FL 33913 Phone 239 337 2116 Toll Free 888 203 7248 Fa...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ ...

Page 24: ...for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water con...

Reviews: