Note: It is recommended to separate the module from the pump and leave the connector PCB attached
to the pump rear. Disconnect the Module ribbon from the pump by use of the ribbon eject levers on the
connector board.
Recommended control cable: metric = 0.14sq mm - 2.5sq mm solid and 0.14sq mm - 1.5sq mm
stranded. USA = 26AWG - 14AWG solid and 26AWG - 16AWG stranded. Cable: circular. Max/min outside
diameter to ensure a seal when passed through the standard gland: 9.5 mm-5 mm.
The cable section
must be circular to ensure a seal
.
Never apply mains power to the terminals within the N module. Apply the
correct signals to the terminals. Limit signals to the maximum values shown.
Do not apply voltage across other terminals. Permanent damage, not covered
by warranty, may result. The maximum rating on the relay contacts of this
pump is 30 V DC; maximum load 30 W.
Note: Also suitable for low power: ie, 1 mA at 5 V DC minimum.
Key to symbols
RUN
STOP
CLOCKWISE ROTATION
ANTI-CLOCKWISE ROTATION
INPUT
OUTPUT
MANUAL (KEYPAD) CONTROL
ANALOGUE (4-20mA/0-10V) CONTROL
EXTERNAL VOLTAGE INPUT
DRY (NO LEAK)
WET (LEAK DETECTED)
KEYPAD DIRECTION CHANGE
REMOTE DIRECTION CHANGE
START DOSE ON RISING EDGE
Connector No.
Function
Input or
output
Configurable
Signal response
J1
AUTO/MAN
AUTO/MAN
No
Standard N Module: Input/output Connectors
32
m-630dun-en-08.2