23.4 313D and 314D flow rates
Flow rates were obtained using silicone tubing with the pumphead rotating clockwise,
pumping water at 20°C with zero suction and delivery pressures. For critical applica-
tions determine flow rates under operating conditions.
Flow rates, 313D, 1.6mm wall (ml/min)
bore
mm
0.5
0.8
1.6
3.2
4.8
6.4
8.0
in
1/50
1/32
1/16
1/8
3/16
1/14
5/16
#
112
13
14
16
25
17
18
15-400 rpm
3-400 rpm
1.5-220 rpm
0.45-12
0.09-12
0.05-6.6
1.1-28
0.21-28
0.11-15
4.1-110
0.81-110
0.41-59
15-400
3.0-400
1.5-220
33-880
6.6-880
3.3-480
54-1400
11-1400
5.4-790
75-2000
15-2000
7.5-1100
Flow rates, 314D, 1.6mm wall (ml/min)
bore
mm
0.5
0.8
1.6
3.2
4.8
6.4
8.0
in
1/50
1/32
1/16
1/8
3/16
1/14
5/16
#
112
13
14
16
25
17
18
15-400 rpm
3-400 rpm
1.5-220 rpm
0.45-12
0.09-12
0.05-6.6
0.90-24
0.18-24
0.09-13
3.8-100
0.75-100
0.38-55
13-340
2.6-340
1.3-190
29-760
5.7-760
2.9-420
45-1200
9.0-1200
4.5-660
60-1600
12-1600
6.0-880
Note: 314D pumpheads should run no faster than 300 rpm when in continuous use.