
4.
Ensure that the valve is firmly in place so that the panel is flush with the outer edge of the
enclosure.
5.
Insert the 4 retaining screws in each corner of the grip panel and tighten them using the 5-
mm hex wrench.
6.
Reconnect the source exhaust by threading the tubing through the hose brackets
alongside the instrument and inserting the tubing into the elbow connector. See
Connecting the internal source exhaust valve (KAD)
Note:
• Routing the exhaust tubing through the hose brackets positions the tubing at a
downward gradient. Doing so prevents the pooling of solvents.
• To prevent the pooling of solvents in a front configuration, ensure that the source
exhaust tubing is positioned forward on a downward gradient.
June 15, 2017, 715003956 Rev. F
Page 58