To perform Span Calibration:

Config Mode   Calibration Menu   Span Calib

 Select Sensor as for Bump Test   Span Calib

• Apply gas to gas inlet, on the top of 

the unit.

• Press Start and wait for the count-

down timer and calibration result.

• Switch off the gas and exit Calib Mode.

Maintenance and Service

For pumped units, replace the inlet filter a 

regular basis, especially when used in high-dust 

environments. For battery or sensor replacement 

and other maintenance or services, please contact 

WatchGas or refer to the User Manual.

V1.23

WatchGas

Sextantstraat 61

2901 ZZ Capelle aan den IJssel

The Netherlands

+31 (0)85 01 87 709

D e t e c t i o n

Summary of Contents for POLI

Page 1: ...Quick Start Guide D e t e c t i o n POLI Multi Gas Detector...

Page 2: ...liser uniquement la batterie WatchGas PN M004 3002 W00 12 01 09 0001 WARNING Never operate the monitor when the cover is removed Remove the monitor cover only in area known to be non hazardous Use onl...

Page 3: ...r ayant une concentration en oxyg ne sup rieure 21 La substitution de composants compromettra l aptitude la s curit intrins que et annulera la garantie Il est recommand de tester avec un gaz de concen...

Page 4: ...icon on the LCD shows the charge status During active charging the icon cycles from empty to full and a yel low alarm light remains on which switches to green when completely charged Charging Comm Ca...

Page 5: ...alarm function Two text and symbol boxes at the LCD bottom are mapped to the Left OK and Right keys The text or symbol in the boxes will vary with the menu Configuration Mode Hold down the Left and R...

Page 6: ...ighlighted then press Left to exit Scrolling Numerical Entries While modifying numerical values such span values or alarm limits the function of the Left OK key can be switched between number increase...

Page 7: ...and alarms functioning without doing a precise calibration To enter bump test Config mode Calibration Menu Bump Test Select sensors to be bump tested All sensors are selected by default Press Right t...

Page 8: ...key during count down The Oxygen sensor is set to 20 9 during Fresh Air calibration and can be zeroed using the Nitrogen Calib menu Span Calibration Span calibration uses known concentrations of gas...

Page 9: ...ion result Switch off the gas and exit Calib Mode Maintenance and Service For pumped units replace the inlet filter a regular basis especially when used in high dust environments For battery or sensor...

Reviews: