7
Single gas detector
PDM
3.4. ALARM SET POINTS
Gas
O
2
CO
H
2
S
H
2
SO
2
NH
3
NO
2
Low alarm
19%
25 ppm
5 ppm
100 ppm
1 ppm
20 ppm
5 ppm
High Alarm
23%
25 ppm
5 ppm
100 ppm
1 ppm
20 ppm
5 ppm
NOTE:
Alarm setpoints can be changed using via WatchGas IR link.
3.5. DEFAULT CALIBRATION CONCENTRATIONS
Gas
O
2
CO
H
2
S
H
2
SO
2
NH
3
NO
2
Concentration
18.0% Vol. 50 ppm
10 ppm
500 ppm
5 ppm
50 ppm
5 ppm
4. EVENT LOG
Last 30 events are stored on the PDM. Once 30 events are stored, the oldest log events get
overwritten. Stored log events can be transferred via WatchGas-IR Link.
Each alarm event is recorded as follows:
• Types of alarms
• Alarm concentration in ppm or %
• Peak concentration
5. CALIBRATION
CAUTION
Initial calibration is performed on all devices prior to shipment. Once received,
calibration should be regularly depending on frequency of use.
5.1. FRESH AIR CALIBRATION
To enter the calibration menu press the button x times untill the calibration concentration is dis-
played. Then hold down the button for 3 seconds. Press and hold the key for 5 seconds to enter
the calibration mode ( ),
icon and ‘CAL’ mark will appear on the LCD.
Press the key for three seconds to initiate calibration. When calibration begins, a countdown
Fresh Air Calibration
Standard Gas Calibration