5

Single gas detector

PDM 

2. ACTIVATION 

In a safe environment, press and hold down the button (

) for three seconds, after the three 

second  countdown the  monitor will turn  on.  Gas type  and firmware version  (eg. 1.2.6) will  be 

displayed. During a 10 seconds countdown, the device will stabilize itself. After stabilization is 

complete, 

 will appear on the display screen and the device will move to the Measuring mode.

In the event that stabilization of the device fails,   will appear on the display and gas measure-

ment mode will not be entered. In this case, perform calibration or contact authorized reseller 

for repair / return information.

CAUTION

Appropriate  calibration  of  the  device  is  required  prior  to  operation.  Always 

ensure that the device makes the proper detection response to the pertinent 

gas. Verify that debris that could interfere with the detection of gas are not 

blocking the area where gas is to be detected.

3. MODE

3.1 MEASURING MODE

When  activated,  in  the  detection  mode,  gas  concentration  or  remaining 

life appears on the screen. Oxygen concentration is displayed in percent 

by volume (%Vol) and toxic concentration is displayed in parts per million 

(ppm). per million (ppm).

3.2 DISPLAY MODE

In the Detection Mode, by pressing push the button for one second, the following icons will        

appear in order: 

 (only for oxygen)

 -> 

 -> CLR -> Remaining Life time(

 ,

 , ) ->Low 

alarm set point ->High alarm set point -> Firmware version -> Calibration Concentration 

Summary of Contents for PDM

Page 1: ...USER MANUAL Single gas detector PDM we watch gas where you can t...

Page 2: ...5 3 3 Alarms Battery Test Failure display 6 3 4 Alarm Set Points 7 3 5 Default Calibration Concentrations 7 4 EVENT LOG 7 5 CALIBRATION 7 5 1 Fresh Air Calibration 7 5 2 Standard Gas Calibration 8 5...

Page 3: ...of the device e g indoors vs outdoors the value of the measured gas concentration can suddenly change Please use the PDM after the gas concentration value has stabilized Severe vibration or shock to...

Page 4: ...R communication port 1 PRODUCT OVERVIEW 3 6 1 2 4 DISPLAY SYMBOLS 5 6 7 Alarm Low Alarm High Alarm Stabilization Success Failure Fresh Air Calibration Standard Gas Calibration Remaining Months Remaini...

Page 5: ...ion CAUTION Appropriate calibration of the device is required prior to operation Always ensure that the device makes the proper detection response to the pertinent gas Verify that debris that could in...

Page 6: ...crease and the alarm will stop Alarm Alarm standard LCD Display Alarm and Vibration 1st Alarm Exceeds 1st alarm Icon concentration 2nd Alarm Exceeds 2nd alarm Icon concentration Remaining Life Below 3...

Page 7: ...a WatchGas IR Link Each alarm event is recorded as follows Types of alarms Alarm concentration in ppm or Peak concentration 5 CALIBRATION CAUTION Initial calibration is performed on all devices prior...

Page 8: ...5 2 STANDARD GAS CALIBRATION Attach the calibration adapter to the PDM and to a gas cylinder with a concentration matching the calibration concentration Check 3 1 Display Mode to check the calibratio...

Page 9: ...ration always verify sensor response and alarm functionality prior to using the detector by applying a known concentration of gas from another gas cylinder 5 3 REMAINING CALIBRATION DAY In the standar...

Page 10: ...ce the bump test interval expires Bts message will flash Put the PDM in the docking station with a valid and full gas cylinder Alternatively attach the calibration adapter to PDM and a valid and full...

Page 11: ...Lithium Primary Battery SB AA02 P 3 6V 1 2Ah MEASUREMENT Diffusion HOUSING Polycarbonate and rubber ACCURACY DEVIATION 2 3 IP RATING IP67 SAFETY CERTIFICATIONS ATEX II 1G Ex ia IIC T4 Ga CSA Class 1...

Page 12: ...S Ex ia IIC T4 KTL 16 KA2BO 0457 INMETRO Ex ia IIC T4 Ga BVC16 5919 02 Compliance Electromagnetic Compatibility Directive 2014 30 EU Standards The electrical apparatus and any acceptable variations to...

Page 13: ...rty s misuse neglect or improper installation testing or calibrations Any unauthorized at tempt to repair or modify the product or any other cause of damage beyond the range of the intended use includ...

Reviews: