Analog Output

 

 

 

 

 

 

 

8.3.4. 

Communication Function

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8.4. Outer Dimensions 

 

 

 

 

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9. 

Timing Chart

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Conversion method: PWM conversion 
Resolution: 13 bits or equivalent 
Scaling: Digital scaling 
Response rate: Approx. 0.5 seconds

* A ripple of 4 to 20 mA is caused at a resistive load of 250 

Ω

 and output of 20 mA.

Output Type

Ripple

0 to 1 V

0 to 10 V

1 to 5 V

4 to 20 mA

±25mVpp

±50mVpp

Resistive Load

10 k

Ω

 or more

10 k

Ω

 or more

10 k

Ω

 or more

550 

Ω

 or less

Accuracy

±(0.5% of FS)

HH

HI

GO

LL

RE

PH

DZ
P1

P2

P3

ME

LO

33

68

80

87

72

7

1

1

8

3

6

*3 + (*1 x 2) + 500 

μs   

*3 + (*1 x 2) + 500 

μs + *4

Internal
sampling 

Set  sampling

S/H terminal

P.C (BCD defined)

*1

*2

*3

Comparison 
output updating

*4

Start/hold type A 

Instruction Manual for MODEL A7X18-X(Temperature Measurement)                                              11/12   

*1 Set sampling

This is the practical sampling rate of the A7000 series that is set using the AVG 
parameter of condition data (see the table below). 

*2 P.C signal output width

This differs depending on the set sampling and is as shown below: 

*3 External start signal 

An external start signal should be 500 

μs to one period of set sampling. 

* RTD: The speed is half of the 
value in the table on the left.
  TC: The speed is quarter of 
the value in the table on the left.

*4 Comparison output delay time 

Relay output: 10 ms maximum, photocoupler output: 200 

μs maximum 

Approx.   4.8  s 

Approx.   1.92 s 

Approx.   96 ms

Approx. 192 ms 

Approx. 384 ms 

Approx. 768 ms 

Approx. 960 ms

50

Set AVG 

Count

20.833 times/sec 

Set Sampling Rate

Approx.    48 ms

Set Sampling Period

100

200

400

800

1000

2000

5000

10.4165 times/sec

5.20825 times/sec 

2.604125 times/sec 

1.3020625 times/sec

1.04165 times/sec

0.520825 times/sec 

0.20833 times/sec

50

100

200

400

800

1000

2000

5000

Approx. 14.4 ms 

Approx. 30 ms 

Set AVG Count

P.C Output Width

Start-stop transmission system

38400bps/19200bps/9600bps/4800bps/2400bps

1bit

7 bits/8 bits

Even parity/odd parity/none

1 bit/2 bits

ASCII codes

Non-procedural

* For more information on the communications function such as the send/receive format and 
commands, see the separate A7000 Communications Function Operation Manual. 

CR+LF/CR

RS-232C

Full duplex

15 m

1

TXD, RXD, SG

Two-line semi-duplex (polling selecting system)

500 m maximum

A maximum of 31 meters

Non-inverse (+), inverse (-)

BCC (block check character) checksum

RS-485

Synchronous System

Communication System

Transmission Rate

Start Bit

Data Length

Error Detection

Stop Bit

Character Code 

Data Communication 

Control Procedure
Signal Name in Use
Number of Units 
Connected

Line Length

Delimiter

Summary of Contents for A7X18-X

Page 1: ...ice 4 Every effort has been made to ensure accuracy in the presentation of this manual However should you have any questions or should any errors or omissions come to your attention please contact you...

Page 2: ...oupler Collector output terminal of LO output photocoupler Collector output terminal of LL output photocoupler Emitter terminal of GO output photocoupler Common emitter terminal for LO and LL output p...

Page 3: ...terminal of HH output photocoupler Start Hold terminal Enabled when it has the same potential as or is shorted with terminal COM Peak Hold terminal Enabled when it has the same potential as or is sho...

Page 4: ...16 32 Large filtering effect Selects a display change width to limit variations in display If this parameter is set to 5 the least significant digit displays only 0 or 5 Selects display color Only wh...

Page 5: ...log output is provided Unit of temperature indication is set Main Setting Target and Precautions Setsthelowlimit of thedisplayablerange Avaluelower thanthedigital limiter LOsetpoint isnot updatedandis...

Page 6: ...ed returning the display to DLHI Condition data Measurement operation Pressing for a prolonged time Comparator data Scaling data Shift data Pressing the Enter key causes the instrument to store data a...

Page 7: ...an example of setting the tolerance judgment type in the next subsection Press the Increment key several times to move to the comparator data menu Press the Mode key to display comparison output type...

Page 8: ...press the Mode key the comparison output type setting content appears O U high and low limit judgment Err tolerance judgment Press the Increment key several times to move to the comparator data menu...

Page 9: ...s have the same potential 6 5 Digital Zero Function The digital zero function regards display at any timing as zero and indicates varying width from this point on This function can be activated or dea...

Page 10: ...measurement PC width depends on sampling Output logic Switchable PC logic not switchable Output signal TTL level fan out 2 COMS compatible Open collector output NPN type Measured data Transistor ON wh...

Page 11: ...he value in the table on the left 4 Comparison output delay time Relay output 10 ms maximum photocoupler output 200 s maximum Approx 4 8 s Approx 1 92 s Approx 96 ms Approx 192 ms Approx 384 ms Approx...

Page 12: ...D converter input or that power ON delay time is valid The zero point of contact compensation circuit is an unstable state temporarily It returns automatically by the sampling of several times Indica...

Reviews: