5.4. Shifting to the Parameter Setting Mode
5.5. Condition Data Setting Procedure
This section describes an example of setting the number of moving
averaging times as a typical example. Take the same steps for other
parameters.
5.6. Scaling Data Setting Procedure
This section describes the concept of analog output scaling and an
example of setting a full-scale indicated value or analog output HI as a
typical example. Take the same steps for other parameters.
* If both AOHI and AOLO are set to the same value, Err6 is caused,
returning the display to AOHI.
* The digital limiter is the function that limits an input value to the
relevant set value even if a signal higher than or lower than the
indicated value set to DLHI or DLLO respectively is input. The
setting condition is DLHI > DLLO, and if this condition is not met,
Err5 is caused, returning the display to DLHI.
Condition
data
Measurement operation
Pressing for a prolonged time
Comparator data
Scaling data
Shift data
Pressing the Enter key causes the instrument to store data
and then return to measurement operation
①
Press the Enter key for approx. 3 seconds during
measurement operation.
②
Press the Mode key to move to the number-of-averaging
times setting.
④
Press the Mode key to move to the number-of-moving
averaging times setting.
⑤
Press the Increment key several times to set a desired
value (this setting example sets the number of moving
averaging times to "4").
Measurement operation
Pressing for a prolonged time
or
③
Press the Increment key to move to moving averaging
setting.
⑥
Press the Enter key to return to measurement operation.
(If the Mode key is pressed, you will be returned to the
setting menu; press the Increment key to move to the
next parameter.)
Instruction Manual for MODEL A7X18-X(Temperature Measurement) 6/12
Display
: 0 to 800
4 to 20mA
:
Output
AOLO : 0
A O H I : 800
Output
Display
20mA
800
4mA
0