3 Safety Guidelines
Configuring and operating this equipment requires
precise knowledge
and
observance
of the instructions in this
user manual
. The equipment is designed only for its intended
application.
WARNING:
Servicing and repairs should be carried out only by authorised specialists.
Disconnect the power plug before starting any maintenance work.
Make sure that the equipment is connected to the correct power source.
The safety valve on the pressure tank opens automatically if the operating
pressure is more than 7 bar.
The pressure can be activated only if the door is closed completely, since a
safety door contact is installed.
Dry-run protection, which switches off at 125 °C. The integrated thermal fuse
works automatically. Please just wait until the unit cooled down.
Warning: Risk of scalding during the Hot Polymerisation.
Mount the thermal protection cover firmly before start-up (see p. 8)
Never open the lid while the unit is under pressure!
Always open the lid slowly!
Display information have to be followed!
Please wear heat-resistant gloves.
For safety-related reasons (hot water, pressure), only the valves and hoses
meant for this purpose must be used for draining. If parts of different
design are used/ installed (self-made), there is a
risk of scalding!