Getting Started Guide: Cobra OCT Spectrometers   

Page 6 

Cobra 1050, Cobra 1300, Cobra 1600  

Camera: SUI GL-2048 

Camera Link Card: DALSA LX1 

To test your Cobra Spectrometer with a National Instruments 

DALSA LX1 

card, make sure that the card 

is installed in the PC and the Cobra spectrometer is connected. To test your SUI camera enabled Cobra 

Spectrometer with Teledyne DALSA LX1 card, make sure the card is installed in the PC and the Cobra 

spectrometer is connected as described in the Setting up the Cobra OCT Spectrometer section. 

Software Prerequisites 

 

DALSA x64 Xcelera-CL version 1.01 or newer  

 

Teledyne DALSA CamExpert 7.40 or newer  

 

Wasatch Photonics Specific Configuration Files 

The following files are required, and are provided on an instrument specific configuration basis with 

your Cobra spectrometer. 

 

COBRA_SensorsUnlimited_2048.ccf 

Look for the calibration information and configuration files for your particular instrument on the USB 

storage drive provided, or contact us for product support with your device serial number if these files are 

needed. 

Sapera Configuration 

Start the Sapera Configuration program and assign the COM 

port for the camera link serial connection as shown. Reboot the 

computer for the serial port changes to take effect. 

Important Note 

Note that no serial configuration over camera link is required, 

as SUI cameras come from the factory pre-configured. They will 

automatically read from the sensor and put data on the camera 

link cable on power up. 

 

 

 

 

 

Summary of Contents for Cobra OCT

Page 1: ...t frequencies or wavelengths to different pixels in the sensor This provides spectral information about the input the light signal The electrical output from camera is transferred to a personal computer using a camera link connection Data Flow It is important to the make right choice of camera link card and software for your application Cobra S spectrometers have the ability to communicate directl...

Page 2: ...alled SDR Shrunk Delta Cables to the spectrometer Note in case the camera is being operated at a rate where medium or full camera link mode is used two camera link cables are required Step 4 Connect the larger end of camera link cable s called MDR Mini D Ribbon to the camera link card installed in your computer Step 5 If the camera is USB powered like the Cobra with a Wasatch internal camera conne...

Page 3: ... for NI MAX to communicate to the camera e2v 2Taps icd Configuring the Camera to Work with NI Max The camera must first be configured to have the same settings as defined by the ICD file linked above To do this launch CommCamm and then follow the prompts for an autodetect If the device is powered and the cameralink cables are properly connected you will see an image similar to the following image ...

Page 4: ...camera file we can now retrieve our data through National Instrument s Measurement and Automation Explorer NI MAX Launch NI MAX and navigate to the connected camera This will look similar to the image below Right click on the connected camera Channel 0 and load the supplied camera file by right clicking selecting Open Camera and then selecting the ICD file ...

Page 5: ...Once done you can now select the Grab button above the image and will now see the test checkerboard pattern begin scrolling in the window similar to the image below You can now go back to CommCam to disable the test image and begin using your spectrometer ...

Page 6: ...ch Photonics Specific Configuration Files The following files are required and are provided on an instrument specific configuration basis with your Cobra spectrometer COBRA_SensorsUnlimited_2048 ccf Look for the calibration information and configuration files for your particular instrument on the USB storage drive provided or contact us for product support with your device serial number if these f...

Page 7: ...ly downloaded Cobra_ model name ccf file under configuration dropdown menu 3 Click the Grab button 4 Click the Graph icon at the far right of the icon bar then choose the line profile option to verify that the spectrometer is returning high gain data 5 The image to the right is the spectral output shown after clicking the Graph Icon ...

Page 8: ...d and are provided on an instrument specific configuration basis with your Cobra spectrometer GL2048L 10A_icd0170rev6 icd Look for the calibration information and configuration files for your particular instrument on the USB storage drive provided or contact us for product support with your device serial number if these files are needed Measurement Automation Explorer 1 Copy the supplied GL2048L 1...

Page 9: ...step 1 above the GL2048L 10A_icd0170rev6 icd file was copied to the NI IMAX icd file directory To select that file for usage in IMAX right click the Channel 0 entry Go to Camera Goodrich GL2048L rev6 file as shown in the image here Note that no serial configuration over camera link is required as SUI cameras come from the factory pre configured They will automatically read from the sensor and put ...

Page 10: ...Getting Started Guide Cobra OCT Spectrometers Page 10 8 Open the TIF file in ImageJ select a line of data in the image click analyze plot profile and you should see something like the following ...

Page 11: ... port with interleaved pixels Control of the spectrometer happens via the serial port on the Camera Link card The following parameter values should be used for the serial communications Baud Rate 9600 Parity None Data bits 8 Stop bits 1 For external triggering the CC1 line on the base Camera Link connector is used Triggering should use LVDS values compatible with the Camera Link interface with tri...

Page 12: ... command can be sent without a parameter to read the current value This should only be set when the camera is NOT scanning i e after an lsc 0 command The integration time must be at least 2 microseconds less than the line time ltm Line Time Command ltm Response ok n Remarks Sets the line time in microseconds Valid range is 22 to 65 535 µs The ltm must be larger than the integration time This comma...

Reviews: