Technical data
MN-111 | Instructions for Use | MN-111 W&H Med ENG Rev12 | 2019/08/31 | © 2019 W&H Sterilization Srl
87
STANDARD STERILIZATION PROGRAMS
The sterilizer offers six preset FDA cleared sterilization programs that comply with the American National Standard ANSI/AAMI ST55:2010:
Note
1
: Mixture of instruments and/or packages within each program may vary. Not to exceed maximum load weight.
CUSTOM STERILIZATION PROGRAMS
WARNING! The custom programs are not FDA cleared and it is the responsibility of the user to validate these programs.
The sterilizer provides the following custom programs. These programs are disabled by default.
Program
Sterilization Temperature
Sterilization Time
(settable)
Drying time
(settable)
Custom A
250 °F (121 °C)
30–60 minutes
30–60 minutes
Custom B
270 °F (132 °C )
4–30 minutes
3–60 minutes
Custom C
273 °F (134 °C)
3–30 minutes
3–60 minutes
STERILIZATION PROGRAM OPTIONS (200–240 V AC VERSION)
WARNING! The sterility of items processed unwrapped is compromised on exposure to non-sterile environments. Ensure that items are
dry when removed from the sterilizer. If the default drying time is not adequate for the load to be sterilized, additional drying time can
be added (see "Sterilization cycle management" on page 50). To load properly the sterilizer see "Prepare the sterilizer" on page 50.
Unwrapped
Wrapped Medium
Load
Wrapped Large
Load
Wrapped XL Load
Low Temperature
Sterilization temperature
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
270 °F (132 °C)
250 °F (121 °C)
Sterilization program type according ANSI/AAMI ST55:2010
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-
removal: SFPP
Dynamic-air-removal:
SFPP
Gravity
displacement