Maintenance
MN-111 | Instructions for Use | MN-111 W&H Med ENG Rev12 | 2019/08/31 | © 2019 W&H Sterilization Srl
69
1200-cycle or yearly maintenance
REPLACING THE DOOR GASKET
Notice: If you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page 65.
Proceed as follows:
1
Open the chamber door.
2
Remove the used door gasket by hand.
3
Carefully clean the seal seat and the inside face of the chamber door.
4
Moisten the new seal with water.
Summary of Contents for Lexa MN-111
Page 1: ...MN 111 Instructions for Use MN 111 W H Med ENG Rev12...