background image

 

 

 

Osmo

 

 

         

 3 

 

INTRODUCTION 

 
 
This manual provides instructions for:  
 

Safe and efficient operation of the equipment 

The correct installation 

Continuous and scheduled maintenance. 

 

 

The dimensions mentioned in this manual are subject to change without notice.  

 

 

Drawings  and  any  other  documents  delivered  with  the  equipment  belong  to  the  manufacturer,  who 
reserves all rights; they cannot be made available to third parties.  

 

 

The reproduction, either total or in part, of the texts and the illustrations in this manual is forbidden.  

 
 
Note:  If the Osmo demineralizer is used to supply sterilizers with demineralized water, it is imperative to 

monitor  the  water  quality.  If  the  supplied  sterilizers  do  not  feature  a  water  conductivity  sensor  to 
monitor the water quality, we recommend using the manual water conductivity meter Primo 5

 
 

GENERAL WARNINGS 

 

 

The equipment should be used in compliance with the procedures described in the manual and never 
for purposes other than those detailed in the manual. 

 

The  user  is  responsible  for  installing,  operating  and  servicing  the  equipment  in  accordance  with  the 
instructions listed in this manual.  

 

Only use genuine manufacturer resin cartridges dedicated for this product. 

 

Do not drink the water produced by Osmo. 

 

Do not block or crush the inlet or outlet pipes of the equipment or the sterilizer. 

 

If the spray gun is not used (manual filling) we recommend closing the spiral tube’s connecting 
tap

 

If  the  equipment  is  not  operated  according  to  the  instructions  in  this  manual  or  is  not  properly 
maintained, the manufacturer cannot be held responsible for any fault, damage or malfunctioning of the 
equipment.  

 

If the water shows high quantities of turbidity, chlorine, iron and manganese, silicon, or a level of TDS 
(Total Dissolved Solids) above 2,000 ppm, the effectiveness and above all the life-span of the osmotic 
membrane will be significantly affected. 

 

Do not use the device with water over 20°C 

 

If the device is not used continuously it may emanate unpleasant smells.

  

 

 
Intended use: This manual is an integral part of the product and must be kept close to the equipment for 
easy and quick reference. The equipment is to be used only for the purpose for which it is designed. 
 

Osmo: Equipment designed for the demineralization of tap water  

Summary of Contents for Osmo

Page 1: ...y it is necessary to read this manual carefully All our products comply with the current safety norms and do not involve any risk for the operator when used according to the instructions provided Whil...

Page 2: ...NAMEPLATE 8 SYSTEM COMPONENTS 8 PRELIMINARY INSTALLATION INSTRUCTIONS 9 INSTALLATION 10 PRIMING 12 EXAMPLES OF USAGE 13 USE AND OPERATION OF THE DEMINERALIZER 14 PRECAUTIONS TO AVOID WATER CONTAMINAT...

Page 3: ...ing and servicing the equipment in accordance with the instructions listed in this manual Only use genuine manufacturer resin cartridges dedicated for this product Do not drink the water produced by O...

Page 4: ...irritation due to the abrasive effect of the granules First aid procedures Skin contact The product cannot be absorbed by the skin Can provoke dehydration or mild irritation due to the abrasive effect...

Page 5: ...case of disputes the date on the delivery note relating to the equipment serial number will be valid The warranty does not cover defects or damage caused by Incorrect installation Inadequate maintenan...

Page 6: ...the package and keep it in case of future needs Open the package and check that the contents correspond to the delivery note there is no evidence of any damage In the event of damage or missing items...

Page 7: ...ube with spray gun 2 9 litre storage tank with a tap theoretic volume 13 19 litres actual volume 9 10 litres 3 Power Supply transformer 240Vac 24 Vdc 4 Water Block valve with filter and setting key wi...

Page 8: ...Resin filter cartridge Pressure control switch Storage tank One way valve Tap optional Spray gun SAT optional Power supply unit Model name Serial number Working pressure Country of manufacture Read t...

Page 9: ...n nearby to dispose the wastewater discarded by the osmotic membrane max 1 5m WB Valve The safety valve Water Block is fitted with a special one way valve which prevents the return of water into the h...

Page 10: ...all the holders A B C Connect the water block group and one way valve to the water network tap Insert the tube into the fitting and tighten with a ring nut Connect the tube to the pump s inlet fittin...

Page 11: ...older Insert the supplied red draining tube on the fitting and tighten with a ring nut Position the tube in a draining sink or a collecting basin A B C Open the devices tap Connect the transformer s 2...

Page 12: ...lug hole Figure 4 Example of installation Connection to Spray gun tap optional SAT optional ATTENTION After setting up the system it takes 5 minutes to produce demineralized water It takes approx 140...

Page 13: ...re 5 Possible uses of the device Manual use with spray gun standard for manual filling of the sterilizer tank and washing of instruments Automatic use to fill the sterilizer using the SAT option Manua...

Page 14: ...ith the pressure control switch set to 3 bar PRECAUTIONS TO AVOID WATER CONTAMINATION CAUTION In accordance with subsection 4 5 of EN1717 in order to avoid water quality degradation and bacterial grow...

Page 15: ...he performance of the product In the event of unexpected contamination of the water coming into the system for example the presence of soil all the cartridges can get damaged Replace the intermediate...

Page 16: ...t the use in unsuitable conditions can drastically reduce this In order to replace the membrane first unplug the power supply and close the water inlet and the storage tank valve Proceed as shown in t...

Page 17: ...of EN1717 have the Water Block and the one way valve checked at least once a year in case a sterilizer is attached to the system this could be done during the scheduled maintenance visits for the ste...

Page 18: ...LOCK CARTRIDGE A813016X 5 RESIN CARTRIDGE A813017X 6 OSMOTIC MEMBRANE 75 GPM 95 98 REJECTION A813018X 7 O RING A813014X 8 PLASTIC KEY FOR CARTRIDGE HOLDER A812101X 9 STORAGE TANK INCLUDING TAP A813007...

Page 19: ...U382016X 16 MANOSTAT AND WIRING KIT A813011X 17 ONE WAY VALVE A812116X 18 UV KIT A813012X 19 KIT WATER TAP A813013X 20 WATER PIPE 5 m 6 35 mm A813020X 21 PRESSURE REDUCER X 3 BAR A813025X 22 CATRIDGE...

Page 20: ...t of the storage tank 3 5 kg Tank nominal volume 19 liters Tank storable volume ca 9 liters Electric power supply 100 240 VAC 50 60 Hz 1 0A with separate voltage transformer Max absorbed power 40 W Cl...

Page 21: ...ened with the aid of a wrench the same must be applied to the appropriately octagonally shaped part of the housing A 2 Screw the supply hose C to the lower end of the WB If the supply hose is equipped...