USE AND OPERATION OF THE DEMINERALIZER
USE OF DEMINERALIZER
Once the system has been correctly installed and connected, the demineralizer is able to immediately
supply purified and demineralized water.
If the demineralized water is used to run a sterilizer, you have the option of manually filling the sterilizer
tank with the use of the water gun, or have it filled automatically by using an external filling device (e.g. Sat)
connected to the Osmo device.
The quantity of water produced by the cartridges depends on the quality and hardness of the city water
supply in the area where you live.
The average time it takes to fill the storage tank is approximately 140 minutes (with the pressure control
switch set to 3 bar).
PRECAUTIONS TO AVOID WATER CONTAMINATION
CAUTION
In accordance with subsection 4.5 of EN1717, in order to avoid water quality degradation and
bacterial growth due to a tailback of water in the tubings, the entire system must be drained
before leaving it unused for a longer period of time.
WARNING
Re-using the system after a long period of non-use
To restart the system when it hasn’t been used for a long time
or after the system has been
emptied, water needs to circulate round the entire hydraulic circuit and the water collected in the
circuit needs to be drained for a few minutes ( see chapter installation, points 6-9).
It is also recommended not to leave the storage tank filled with demineralised water for a long
period of time
.
IDENTIFICATION OF THE CARTRIDGES
Mechanical filter
Active carbon filter
Resin filter
Osmotic membrane