19. Servicing
Regular checking of Implantmed and accessories
Regular servicing of function and safety including the accessories is necessary and should be carried
out at least once every three years, unless shorter intervals are prescribed by law. The inspection
must be undertaken by a qualified organization and must include the following procedures:
> Visual inspection for outside damage
> Measurement of device leakage current
> Measurement of patient leakage current
> Visual inspection of internal components on suspicion of safety interference,
e. g. mechanical damage of the enclosure or indications of overheating
> Visually inspecting the ESD spring contact on the underside of the foot control (electrostatic discharge)
> Function test with check to see if the maximum speed can be reached
We recommend that only an authorized W&H service partner (see page 52) should undertake this servicing and checking.