38
Hygiene and maintenance
Motor with cable
Sterilization and storage
W&H recommends sterilization according to EN 13060, class B
Other sterilization methods may reduce the life span of your motor.
For USA and Canada:
Hospital grade sterilization with pre and post vacuum cycle.
> Observe the device manufacturer’s instructions.
> Clean and disinfect prior to sterilization.
> Wrap the motor with cable and accessories in sterile goods packaging according to EN 868-5.
For USA and Canada:
Place the motor and the accessory in a standard handpiece sterilization pouch.
> Make sure that you only remove dry sterile goods from the sterilizer.
> Store sterile goods dust-free and dry.