9
Safety notes
Danger zones M and G
In accordance with IEC 60601-1 / UL 60601-1, the control unit is not suitable for use in potentially explosive
atmospheres or with potentially explosive mixtures of anaesthetic substances containing oxygen or nitrous oxide.
Control unit
The control unit is classed as "conventional equipment" (closed equipment without protection against the ingress of
water).
The OsseoSet
TM
200 Programs 4 and 5 are only designed to be used with the contra-angle handpieces
WI-75 E/KM and WS-75 E/KM approved by W&H. Use of other contra-angle handpieces may result in deviation
from the indicated torques and is under the user’s responsibility. The specified transmission ratios for the
Programs 1 to 5 must always be considered.
Motor and foot control
The motor and foot control are approved for use in zone M (AP).
Please note that at low speeds, it is more difficult to determine that the motor is running.