26
Control unit operation – Changing the torque (P4 – P5)
Adjustable range from 5 up to 50 Ncm. Keeping PLUS / MINUS depressed activates the repeat function and the
values are continuously increased / decreased.
When changing from 5 to 50 Ncm and 50 to 5 Ncm, a longer confirmation signal is heard.
When the adjustable torque in the the forward/reverse drive mode is reached, the motor switches off
automatically.
➊
Press Program button (P4, P5)
➋
Increase torque in 5-Ncm-steps
➋
Decrease torque in 5-Ncm-steps