11
Safety notes
Coolant
The OsseoSet
TM
200 is designed for use with physiological saline solution. Use only suitable irrigation fluids and
comply with the medical data and instructions from the manufacturer. Coolant bottle, coolant bag and sterile
fittings can be obtained from a pharmacist.
Rotational energy
Fast deceleration of the bur can, at times, cause the selected torque to be overloaded as a result of the rotational
energy stored in the drive system.
We therefore advise compliance with the individual manufacturer's instructions for use, particularly when adjusting
superstructure screws.
For safety reasons, these should be screwed in very carefully by hand, or by using a torque wrench.
We would point out that adjusting these screws with an electric motor represents a potential risk as described
above.