41
LED display
Description
Solution
The start button LED
is blue.
Problem with air supply.
> Restore the air supply.
The LED for the right or left
chamber is red.
The start button LED is green.
The maintenance process did not
complete successfully.
> Press the »Start« button.
> Repeat the maintenance process.
The LED for the right or left
chamber is red.
The start button LED is blue.
The maintenance process did not complete
successfully – problem with air supply.
> Restore the air supply.
> Press the »Start« button.
> Repeat the maintenance process.