The cartridges must be at room temperature when they are inserted.
The cartridge spike in the medical device is sharp.
Be careful not to injure yourself on the projecting spike.
Remove the cartridge
compartment cover.
Inserting a cartridge.
> Push the cartridge in until
the limit stop is reached.
e
Removing a cartridge.
> Pull out the cartridge.
Replace the cartridge
compartment cover.
Whenever you replace a cartridge:
>
always clean the cartridge compartment,
>
perform the initial filling.
Observe the manufacturer’s instructions on the use of disinfectant.