Analyzer Initial Start-Up
29
Waltron User Manual
3051 Hardness Analyzer
4
ANALYZER INITIAL START-UP
Before proceeding with analyzer start-up it is absolutely
necessary to check that all procedures for a proper
installation and reagent preparation have been made.
Please verify that all the suggestions and recommendations
have been respected.
4.1
PREPARING THE ANALYZER FOR START-UP
Once installation is complete, proceed as follows to prepare the analyzer for online operation :
1.
Connect the sample line inlet tubing (or filtered sample outlet coming from optional
filtration system) to the fast-loop reservoir installed on the left side of the analyzer
2.
Connect the drain fitting of the fast-loop reservoir to the waste line
3.
Put the reagent inlet tubings into the appropriate reagent containers
a.
Reagent 1 = Cresolphthalein Reagent
b.
Reagent 2 = Ethanolamine Reagent
4.
Put the calibration solution inlet tubing into the calibration solution positioned beneath
the analyzer
5.
Connect the analyzer drain tubing to waste drain line
WARNING!
The drain of the analyzer must be at ambient pressure with
no restriction. Please verify that this condition has been
strictly respected during initial start-up procedures.
6.
Check sample presence in fast-loop reservoir and adjust the sample flowrate so that there
is complete reservoir discharge when a finger is covering the hole at the top of the
stainless steel drain line
7.
Turn the analyzer on using the switch in the electronics compartment
8.
Perform reagents prime (see Section 6.1)
9.
Perform calibration (see Section 6.2 & 6.3)
10.
Set the analyzer to online mode (see Section 6.5)
Summary of Contents for AQUALERT Aqualyzer 3051
Page 52: ...Appendix A Panel Mount Schematics 51 Waltron User Manual 3051 Hardness Analyzer ...
Page 53: ...Appendix B Help Menu 52 Waltron User Manual 3051 Hardness Analyzer APPENDIX B HELP MENU ...
Page 54: ...Appendix B Help Menu 53 Waltron User Manual 3051 Hardness Analyzer ...
Page 55: ...Appendix B Help Menu 54 Waltron User Manual 3051 Hardness Analyzer ...
Page 56: ...Appendix B Help Menu 55 Waltron User Manual 3051 Hardness Analyzer ...
Page 57: ...Appendix B Help Menu 56 Waltron User Manual 3051 Hardness Analyzer ...
Page 58: ...Appendix B Help Menu 57 Waltron User Manual 3051 Hardness Analyzer ...
Page 59: ...Appendix B Help Menu 58 Waltron User Manual 3051 Hardness Analyzer ...
Page 60: ...Appendix B Help Menu 59 Waltron User Manual 3051 Hardness Analyzer ...
Page 61: ...Appendix B Help Menu 60 Waltron User Manual 3051 Hardness Analyzer ...
Page 62: ...Appendix B Help Menu 61 Waltron User Manual 3051 Hardness Analyzer ...