Installation
25
Waltron User Manual
3051 Hardness Analyzer
WARNING!
The sample drain of the analyzer must be at ambient
pressure with no restriction or counter pressure. Please
verify that this condition has been strictly respected during
installation.
Clearance requirements for the analyzer should be about 8” (20 cm) on either side and
about 40” (100 cm) in the front.
Sufficient space for the reagent containers should be provided on the side or beneath the
analyzer.
The reagent containers should be positioned in the reagent bracket or a suitable basin in
case of spills.
NOTE:
Respect the maximum height of
16” (40 cm) between the
reagent's bottle(s) and the reagent's pump(s).
3.5
MOUTING SCHEMATICS
Figure 3.1: The dimensions for mounting the analyzer.
10.16” (258 mm)
9.76
” (248 mm)
16.22
” (412 mm)
14.96
” (380 mm)
1
8
.8
6
” (
479
mm)
2
3
.7
4
” (
603
mm)
Summary of Contents for AQUALERT Aqualyzer 3051
Page 52: ...Appendix A Panel Mount Schematics 51 Waltron User Manual 3051 Hardness Analyzer ...
Page 53: ...Appendix B Help Menu 52 Waltron User Manual 3051 Hardness Analyzer APPENDIX B HELP MENU ...
Page 54: ...Appendix B Help Menu 53 Waltron User Manual 3051 Hardness Analyzer ...
Page 55: ...Appendix B Help Menu 54 Waltron User Manual 3051 Hardness Analyzer ...
Page 56: ...Appendix B Help Menu 55 Waltron User Manual 3051 Hardness Analyzer ...
Page 57: ...Appendix B Help Menu 56 Waltron User Manual 3051 Hardness Analyzer ...
Page 58: ...Appendix B Help Menu 57 Waltron User Manual 3051 Hardness Analyzer ...
Page 59: ...Appendix B Help Menu 58 Waltron User Manual 3051 Hardness Analyzer ...
Page 60: ...Appendix B Help Menu 59 Waltron User Manual 3051 Hardness Analyzer ...
Page 61: ...Appendix B Help Menu 60 Waltron User Manual 3051 Hardness Analyzer ...
Page 62: ...Appendix B Help Menu 61 Waltron User Manual 3051 Hardness Analyzer ...