Troubleshooting and Repair
42
Waltron User Manual
9065 Luminescent Dissolved Oxygen Analyzer
8.3
REPAIR AND MAINTENANCE
Before performing any maintenance on the 9065, make sure
the instrument is disconnected from the power supply during
assembly and disassembly so that there is no risk of an
electric shock.
8.3.1
REPLACING THE COATING HOLDER
If the coating is faulty or damaged, then a new coating holder must be placed on the sensor. The
instructions for replacing the coating are given below.
1.
Remove the sensor from the mounting plate and place it on a level surface so that it cannot roll
onto the floor.
2.
Use a 10mm open or closed ended spanner to
unscrew the coating holder from the measuring
head.
3.
Use a brush or a small screwdriver to remove any
glue or dirt from the screw thread around the light
guide. Do not damage the light guide.
Light guide
Screw thread