User Interface
28
Waltron User Manual
9065 Luminescent Dissolved Oxygen Analyzer
Table 4.2: Configurable settings for the analog outputs.
Item
Setting Options
Default Setting
Configure output 1 O
2
sensor 1, O
2
sensor 2, Temp. Sensor 1,
Temp. Sensor 2
O
2
sensor
Configure output 2 O
2
sensor 1, O
2
sensor 2, Temp. Sensor 1,
Temp. Sensor 2
Temp. sensor
Item
Range
Default
Unit
Minimum value (if O
2
sensor)
0 …. 45,000
0
ppb; ppm; mg/l; ug/l;
% as; %
Maximum value (if O
2
sensor)
0 …. 45,000
2,000
ppb; ppm; mg/l; ug/l;
% as; %
Minimum value (if
temperature sensor)
-
10 … 150
0
° C
° F
Maximum value (if
temperature sensor)
-
10 … 150
100
° C
° F
The settings can be set/changed as follows:
1.
Select ‘Analogue output’ in the Configuration menu and press the Enter button. The
Analog outputs menu will be displayed.
2.
In this menu, use the cursor keys to select the desired item and press the Enter button to
confirm the selection.
3.
Use the cursor keys to select the setting concerned.
4.
Press the Enter button to return to the Analog outputs menu.
5.
Use the cursor keys to select the Return sign and press the Enter button to return to the
Configuration menu.
NOTE:
The settings output 1 and 2 are defined for the measuring
values at the IO. See Section 3.6.5 for the IO connections.
4.5.10
ALARM SETTINGS
In this menu, the alarm level can be set to the desired level. The settings apply to both Sensor 1
and Sensor 2. If only one sensor is connected, then the other choices will not be displayed. An
overview of the possible settings and the default settings is given below.
The settings can be made as follow:
1.
Select ‘Alarm settings’ in the Configuration menu and press the Enter button. The Alarm
settings menu will be displayed.
2.
In this menu, use the cursor keys to select the desired item and press the Enter button to
confirm the selection.
3.
Use the cursor keys to select the setting concerned.
4.
Press the Enter button to return to the Alarm settings menu.
5.
Use the cursor keys to select the Return sign and press the Enter button to return to the