Overview
9
Waltron User Manual
9065 Luminescent Dissolved Oxygen Analyzer
1
OVERVIEW
1.1
TECHNICAL SPECIFICATIONS
Range
[ppb]
0.0 to 2,000
[µg/l]
0.0 to 2,000
[ppm]
0.00 to 2.00
[%] O
2
0.000 to 4.180
[% a.s.]
0.00 to 20.0
[mg/l]
0.00 to 2.00
Temperature
[
o
C]
- 5.0 to 45.0
[
o
F]
23.0 to 113.0
Accuracy *
Measurement
[ppb]
± (1 ppb + 2% of the measured value)
[μg/l]
± (1 μg/l + 2% of the measured value)
[ppm]
± (0.001 ppm + 2% of the measured value)
[mg/l]
± (0.001 mg/l + 2% of the measured value)
Temperature
[° C]
± 0.1° C
[° F]
± 0.2° F
Max. Pressure
[barg]
10.00 barg
[psi]
145.0 psi
[kg/cm
2
]
10.20 kg/cm
2
[kPa]
1,000 kPa
Measurement
frequency
[s]
Adjustable to every 2 sec. up to
999 s + automode
Memory
500 measurements/ O2 probe
Voltage
Standard
90
– 240 V ~ 50 – 60 Hz, 25 W
Optional
24 V DC, 25 W
Installation
Sensor
One or Two Sensor Configurations
Transmitter
Standard
Wall-mounted
Optional
Panel-mounted
IP class
Sensor
IP67
Transmitter
IP67
Dimensions
Sensor
mm
Ø 84 x 240, excluding plug