Overview
10
Waltron User Manual
9065 Luminescent Dissolved Oxygen Analyzer
Transmitter
mm
W178 x H278 x D90
Weight
Sensor
kg
2.2 kg
Transmitter
kg
1.2 kg
O
2
-probes per
Transmitter
2 max.
* The accuracy from the 9065 is determined at 20° C
1.2
INTENDED USE
The Waltron 9065 Dissolved Oxygen Analyzer utilizes new luminescent technology for
measuring dissolved oxygen in water at ppb levels. Luminescent technology has unique features
and benefits compared to traditional dissolved oxygen sensing technologies. The 9065 provides
high accuracy with excellent long-term stability. The 9065 Dissolved Oxygen Analyzer can be
used in a variety of online analysis applications throughout many different industries.
The 9065 must be connected to a sample line. The connections to the flowcell are ¼” (6mm)
Swagelok fittings. The dissolved oxygen must be measured in a full, flowing sample line. Using
the sensor in a pipe that is not full or taking the measurement while the product is not flowing
will cause errors or will it cause damage to the device.
Features:
Analysis range: 0.10ppb – 2000ppb (others available upon request)
No calibration for up to 2 years
Extremely fast response
No sample interference
Excellent results regardless of sample flowrate
Use in liquid and gas applications
High temperature alarm
Multiple sensor configuration
Benefits:
No sensor maintenance (no membrane, no electrolyte)
Simple operation
Compact design