3.0 PROBES
One of the key features of the Wahl Thermocouple Heat-Prober is a standard miniature
thermocouple connector used in the meter. Thermocouple connections are color
coded by type. Use probes with the same color male plug as the female. Each Heat-
Prober is designed for use with a specific thermocouple Type K or S. A thermocouple
of the same type with a standard male thermocouple connector will fit the meter. No
adapters are needed. The thermocouple is compensated where the cold junction is
made and cold junction compensation occurs automatically.
Each meter face indicates the thermocouple type, the temperature range of use, and
the calibration in °F or °C, as shown in Figure 3-1.
THERMOCOUPLE TYPE
RANGE
MODEL-NO.
Type K (Yellow)
-100°F to 2500°F
(-70°C to 1370°C)
2500M
2500MX
2500MVX
1370M
1370MX
1370MVX
Type S (Green)
77°F to 3200°F
(25°C to 1770°C)
3200M
3200MX
3200MVX
1770M
1770MX
1770MVX
Figure 3-1. Heat-Prober Thermocouple Types and Models
Summary of Contents for 2500MX
Page 8: ...4 Figure 2 1 Digital Heat Prober Major Features...
Page 14: ...10 Figure 3 2 Standard Wahl Probes Figure 3 3 Extension Probe Handle...
Page 16: ...12 Figure 3 4 Molten Metal Dipstick...
Page 28: ...24 Figure 5 2 Heat Prober with Wahl Calibrator...
Page 29: ...25 TP2 R29 R9 R20 TP1 R22 TP3 U 3 Figure 5 3 LED Adjustment Points...
Page 30: ...26 TP2 R29 R9 R20 TP1 R22 TP3 U 3 Figure 5 4 LCD Adjustment Points...
Page 32: ...28 Figure 6 1 Program Flow Chart...
Page 43: ...39 Figure 8 1 Block Diagram...
Page 44: ...40 Figure 8 2 LED Component Layout...
Page 45: ...41 Figure 8 3 LCD Component Layout...
Page 54: ...50 Figure 11 1 2500M Accessories and Special Purpose Probes...