Chapter 5, Using Vitals Mobile-TC (TC-100 Controller): Valve Operation
E.H. Wachs
Part No. 79-MAN-04, Rev. I
59
Figure 5-33. Example
Transducer Error
screen. If you
tap
Yes
, the valve operation will continue. If you tap
No
, the valve operation will stop.
Figure 5-34. If you selected
Yes
to continue (left), the
valve operation will continue. If you selected
No
, the
valve operation is stopped. The message
TRANSDUCER
ERROR
is noted on the screen.
Using the valve operator with transducer errors (by clicking
Yes
) is
not
recommended. If necessary, you can still operate
it, but torque control will not be reliable. Use caution to
avoid damaging the valve.