Chapter 8, Troubleshooting Guide:
E.H. Wachs
Part No. 79-MAN-04, Rev. I
197
Use HC-100 stylus.
1. Wider tip stylus will actually work better the a fine tip
stylus because the wider tip will activate more contact
points at a time.
Use Hold To Zoom Feature
1. Press and hold down P1 button
2. Tap on the screen. Area around the tap location will
zoom in allowing finer user touch input.
3. Release P1 button to zoom out
Soft Reset
1. Try soft resetting the device by pressing and holding the
Power (On/Off) key for 5-10 seconds until the Power
Button menu appears. Tap Reset.
Symptom: HC-100 turning off and rebooting
Low battery
1.
Charge battery
2.
Keep HC-100 plugged-in (vehicle DC charger or
AC charger) when not in use
Symptom: Vitals Mobile Controller Stops responding
Faulty machine connection
1. Soft reset HC-100
2.
Check connection with Vitals Diagnostics
•
Vitals Diagnostics will not connect to machine
with faulty connection. Vitals Mobile accepts an
intermittent connection problem.
i. Diagnose source of faulty connection: control
cable, wiring, connector corrosion, faulty equip
-
ment ground, faulty control board.
Summary of Contents for Vitals
Page 2: ...Vitals Reference Manual Part No 79 MAN 04 Rev I E H Wachs...
Page 14: ...Vitals Reference Manual 8 Part No 79 MAN 04 Rev I E H Wachs...
Page 32: ...Vitals Reference Manual 26 Part No 79 MAN 04 Rev I E H Wachs...
Page 80: ...Vitals Reference Manual 74 Part No 79 MAN 04 Rev I E H Wachs...
Page 138: ...Vitals Reference Manual 132 Part No 79 MAN 04 Rev I E H Wachs...
Page 198: ...Vitals Reference Manual 192 Part No 79 MAN 04 Rev I E H Wachs...
Page 224: ...Vitals Reference Manual 218 Part No 79 MAN 04 Rev I E H Wachs...