Each blink will take one second, with half of the second interval being
ON (illuminated) and the other half of the second interval being OFF.
After the sequence concludes, both the Low Battery LED and the
Solenoid Engaged LED will flash four times ON/OFF quickly to
indicate a return to normal display.
3. BUS Injection Test
The user must press and hold the Press-to-Test button for >= 2
seconds before injection occurs. After the Press-to-Test button is
pressed for >30 seconds, the solenoid will turn off. This will avoid a
button or button signal becoming stuck in an active condition.
During a Press-to-Test injection, if either RTD #1 or RTD #2 reads a
temperature less than 200C below BUS set point and it is determined
to be a valid input, then injection will terminate immediately.
4. BUS Set point “Validation” Feedback
Whenever the BUS set point or working fluid changes, the BUS
acknowledges the set point acceptance by causing the Low Battery
LED and Solenoid Engaged LED to blink three times ON/OFF
quickly. This does not imply that the BUS error-checked the values,
other than that the value received was within the range of the BUS
working correctly with prior programming.
Note
The solenoid will not engage if lid is open.
s
Caution
.The back-up system is designed to inject refrigerant (CO
2
or
LN
2
). In the unlikely event of both probes failing, the back-up system will
operate in PWM or Non-PWM mode based on jumper setting (Figure 7-
7)
s
For Hysteresis & Dual probe fault mode settings (Figure 7-7), locate
Jumper J17 near to solenoid/ sensor connector (J1), refer to table below for
operation details:
7-6
Chest Freezer
VWR International
Section 7
Factory Installed Options
Jumper PIN
Status
Hysteresis
Dual Probe Fault Mode
J17 PIN 1 & PIN 2
Short Together/ Jumper Cap Installed
5°C
Non-PWM Mode
J17 PIN 1 & PIN 2
Open/No-Jumper Installed
9°C
PWM Mode
Table 7-2.
Probe Fault Modes
BUS Control Panel
(cont)
Summary of Contents for VWRC0386A
Page 55: ...8 1 VWR Collection VWR International Section 8 Parts ...
Page 56: ...VWR Collection 8 2 VWR International Section 8 Parts ...
Page 59: ...8 5 VWR Collection VWR International Section 8 Parts ...
Page 60: ...VWR Collection 8 6 VWR International Section 7 Specifications ...
Page 61: ...VWR Collection 8 7 VWR International Section 8 Parts ...
Page 68: ...9 1 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 69: ...VWR Collection 9 2 VWR International Section 9 Refirgeration Schematics ...
Page 70: ...VWR Collection 9 3 VWR International Section 9 Refirgeration Schematics ...
Page 71: ...9 4 VWR Collection VWR International Section 9 Refrigeration Schematics ...
Page 72: ...10 1 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 73: ...VWR Collection 10 2 VWR International Section 10 Electrical Schematics ...
Page 74: ...10 3 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 75: ...VWR Collection 10 4 VWR International Section 10 Electrical Schematics ...
Page 76: ...VWR Collection 10 5 VWR International Section 9 Section title ...
Page 77: ...10 6 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 78: ...VWR Collection 10 7 VWR International Section 10 Electrical Schematics ...
Page 79: ...10 8 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 80: ...VWR Collection 10 9 VWR International Section 10 Electrical Schematics ...
Page 81: ...10 10 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 82: ...VWR Collection 10 11 VWR International Section 10 Electrical Schematics ...
Page 83: ...10 12 VWR Collection VWR International Section 10 Electrical Schematics ...
Page 93: ...VWR International D 2 Chest Freezer ...
Page 94: ......