Installation procedures
Mains connection
5-2 18L personal incubator
VWR
Connection to the power supply source
1. Before connecting the device to the power source, check to see if the power supply voltage
corresponds with the specifications on the nameplate on the rear of the device. If the voltage (V)
and current (A) ratings given are not as required, do not connect the device to the power source!
2. Connect the IEC connector to the socket on the rear of the device.
3. Insert the earthed plug [3] of the mains cable in a correctly earthed and fused socket.
4. Make sure the power cord is not subjected to tensile or compressive force.
Power sockets
The socket which the device is connected to is to be fully accessible
when operating the device so as to ensure that the plug can be re-
moved should danger occur!
Condensation
The device should be in the room for a sufficiently long duration be-
fore it is switched on, in order to ensure that it can reach the ambi-
ent temperature. This prevents condensation forming on live
components.