![VWR 620-2675 Operating Manual Download Page 19](http://html1.mh-extra.com/html/vwr/620-2675/620-2675_operating-manual_1057636019.webp)
TD310
Error and system messages | 9
B-H86.0.04.DB201-1.0
19 / 24
9 Error and system messages
Display
Meaning
Possible
causes
Remedy
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range ex-
ceeded
Measurement too
high
Measuring probe or
product defect
The measurement is above
the permissible range
Check measuring probe
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Measuring probe or
product defect
The measurement is below
the permissible range
Check measuring probe
Send in for repair
ERR.7
System error
Error in the product
Switch the product on/off
Replace batteries
Send in for repair