7 | Bases for measurement
TD310
16 / 24
B-H86.0.04.DB201-1.0
7 Bases for measurement
7.1 Sensor/device accuracy
The product has high system accuracy. This is achieved by calibrating the temperat-
ure sensor with the product when it is manufactured. This is achieved with precise
measuring technology and high-quality Pt1000 sensors.
7.2 Possible measuring errors
7.2.1 Immersion depth
Liquids
Immerse to a depth of at least 20 mm and then stir. Otherwise, measuring errors can
occur due to the heat transmission of the sensor tube if the immersion depth is too
shallow.
7.2.2 Limit values
Temperature range
CAUTION
Destruction of the measuring sensor!
When conducting measurements in media with high or very low temperatures, there is
a risk that the measuring sensor is not designed for such extremes.
– The limit values must be observed!