TD310
Operation | 6
B-H86.0.04.DB201-1.0
15 / 24
Representation
_
Parameter
Values
Meaning
_
Zero point correction
OFFS
OFF
No zero point correction
-2.5 .. 2.5
Zero point correction in °C. and/or at °F -4.50 ..
4.50
_
Gradient correction of the temperature
S(AL
OFF
No gradient correction
-5.00 .. 5.00
Gradient correction in %
_
Formula
_
Gradient correction °C:
Display = (measured value –
OFFS
) * (1 +
S(AL
/ 100)
_
Gradient correction °F:
Display = (measured value – 32 °F –
OFFS
) * (1 +
S(AL
/ 100) + 32 °F
_
Outcome of an action
The changed value is saved and the
Adjustment
menu is closed by pressing the
On/
Off button
.
NOTE
The
Adjustment
menu is cancelled if no button is pressed for 2 minutes. Changes
made prior to that are not saved.