![VWR International INCU-Line 150R Instruction Manual Download Page 23](http://html1.mh-extra.com/html/vwr-international/incu-line-150r/incu-line-150r_instruction-manual_1057669023.webp)
Instruction Manual INCU-Line Cooled
Page
23
/
32
Choose RESET and confirm with ENTER
to zero the
recordings.
11 ALARMS
Every alarm is signaled by red diode placed on front panel, sound and
'ALARM'
text shows on display. Pressing
the accept button you can
see date and time of alarm event and kind of alarm. The alarm will appear
when:
1. temperature sensor is damaged
– the alarm message - 'sensor error'.
2. temperature in a given temperature segment, after the set temperature is reached:
a) is higher than the set point at least by the value set by the user in the menu: SETUP / ALARMS /
ALARMTEMP.HI - the alarm message 'HI temp'.
b) is lower than the set point at least by the value set by the user in the menu: SETUP / ALARMS /
ALARMTEMP.LO - the alarm message 'LO temp'.
3. power failure - alarm message 'power failure' only if it was turned on by the user in the menu: SETUP /
ALARMS / POWER ALARM.
4. door has been opened - alarm message
'open door'
will appear after the set time in the menu: SETUP
/ ALARMS / ALARM DOOR.
5. preset time in the user menu has elapsed: SETUP / ALARMS / CLOCK ALARM
– the alarm message
‘time alarm’.
6. when the temperature of protection class 3.3 will be :
a) higher than the set temperature by the user in the menu: SETUP / PROTECTION / HI protection
–
the alarm message ‘HI prot’.
b) lower than the set temperature by the user in the menu: SETUP / PROTECTION / LO protection
–
the alarm message ‘LO prot’.
When the cause of the alarm stops, the alarm may be cancelled. To cancel the information of the
event, press down the enter button
for a while.
Sort of event
Duration
of sound
alarm
Red di-
ode
Duration of break
between sounds
[second]
Terminate of event
Displayed
command
Temperature
sensor fault
0.1s
yes
1s
Replacing of tempera-
ture sensor
SENSOR
ERROR
Time alarm
1s
no
1s
Pressing of any button
TIME ALARM
Power alarm
0.1s
no
5s
Pressing EXIT button
POWER
ALARM
Open door alarm
1s
no
1s
Closing door or press-
ing any button
DOOR OPEN