Instruction Manual INCU-Line Cooled
Page
20
/
32
8 TEST RESULTS MEMORY
In MENU choose LOGGER
To choose from DATA or EVENTS.
In the menu you can set the interval time of saving temperature in the
chamber. Note <00:15> means, that every 15 minutes data of tempera-
ture will be registered in internal memory (max 2048 results). If the
memory will be fulfilled, the data will be over written.
Erase the data.
Reading the data from logger is possible by using a PC application
–
Basic Control
Event window opens the internal event logger.
Command ‘view will display events like: start, stop or end of program
as well as loss and return of power, alarm of over temperature, door
openings.
Sample of displayed events and occur time of the event. After pressing
and pushing
the date of the event will be displayed.
Erasing the event from the memory (max. numbers of saved events is
110), If the memory will be fulfilled, the data will be over written.
Reading the data from logger is possible by using a PC application
–
Basic Control.