VWR Cooled Incubators 

 

 

INCU-Line 68R: 390-0728

 

INCU-Line 150R: 390-0729 
INCU-Line 250R: 390-0731

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 

 

 

 

Before using the equipment, please read carefully 

this instruction manual!

 

 

 

Version 3.6 

 

Summary of Contents for INCU-Line 150R

Page 1: ...C CT T TI I IO O ON N N M M MA A AN N NU U UA A AL L L VWR Cooled Incubators INCU Line 68R 390 0728 INCU Line 150R 390 0729 INCU Line 250R 390 0731 Before using the equipment please read carefully thi...

Page 2: ...Instruction Manual INCU Line Cooled Page 2 32 Legal Adress of Manufacturer Country of Origin Poland...

Page 3: ...Instruction Manual INCU Line Cooled Page 3 32...

Page 4: ...Instruction Manual INCU Line Cooled Page 4 32...

Page 5: ...THE CONTROLLER 15 6 1 Simple mode 15 6 1 1 To stop the unit in the simple mode 15 6 1 2 To start the unit in the simple mode 15 6 2 Switching between simple and complex modes 16 6 3 Complex mode 16 6...

Page 6: ...32 14 1 Housing cleaning 26 14 2 Interior cleaning 27 15 RESUMING OPERATION AFTER A LONGER PERIOD OF TIME 27 16 TROUBLESHOOTING 28 16 1 REPLACEMANT OF FUSE 29 17 PRODUCT IDENTIFICATION LABEL 29 18 WAR...

Page 7: ...tric shocks unplug the power cable holding the protective cover and not the cable itself disconnect the unit from the mains before undertaking any repairs or maintenance works protect the power cable...

Page 8: ...lative humidity of the ambient air to 60 The unit has not been designed to work in highly dusty environments The place of installation should be appropriately ventilated according to its size The unit...

Page 9: ...lots on the other wall To remove a shelf perform the above steps in reverse order To remove the shelf holder from the slots lift it up and slide it towards the rear of the chamber 3 2 Remarks on the t...

Page 10: ...3 of the space below the shelves the samples should be placed in such a way that so that the horizontal surface between the containers does not exceed 1 3 of the width and height of the empty shelf th...

Page 11: ...ectronic controller with LCD display 3 Chamber fan 4 Access port 30mm 5 Shelf 6 Key lock 7 Main switch 8 Internal electrical socket 9 Internal lighting LED 10 Temperature sensor 11 Height adjustable f...

Page 12: ...keys Confirm using the enter button The program continues to set the next parameter If the parameter value is incorrect us ing the selection keys up down select the parameter again and set the correc...

Page 13: ...controlling the operation of the device depend on the type of the device The most advanced software allows storing up to 3 six segment pro grams Simple mode Temperature Time Tp To Tp set temperature...

Page 14: ...after connecting the unit to the mains please switch it on using the on off switch please press down the button which is located below the control panel after a successful start up Switching off the...

Page 15: ...e UP or DOWN to set a new temperature Press ENTER to confirm Use UP or DOWN to set the decimal value and confirm with ENTER A new temperature has been set You can press ESC at any time to cancel 6 1 1...

Page 16: ...its duration To switch from simple to complex mode perform the following steps In the main window press ESC to enter the MENU Enter SETUP using ENTER Press down and choose PROG MODE Press ENTER The s...

Page 17: ...for segment 1 Use and to set a new value then confirm with ENTER Set the decimal value in the same way Now you can set the segment time in dd hh mm format If the program has more than one segment you...

Page 18: ...start a saved program In the main window press Using and choose the program you would like to start and press 6 3 3 To stop a program In the main window press twice Confirm with 7 SETTINGS OF DEVICE...

Page 19: ...protocol for printers with serial port terminal service service and text protocol dedicated software After selecting a protocol with buttons the user is able to de termine its settings Print interval...

Page 20: ...ng the data from logger is possible by using a PC application Basic Control Event window opens the internal event logger Command view will display events like start stop or end of program as well as l...

Page 21: ...NTROL During the first connection PC will automatically install the driver support 4 In program BASIC CONTROL select CONNECT and specify COM port to which the device is con nected By clicking ACCEPT t...

Page 22: ...r the next segment When next set temperature is reached the recording starts again For multi segment programs there are available statistics for each segment If the program has not reached a seg ment...

Page 23: ...S POWER ALARM 4 door has been opened alarm message open door will appear after the set time in the menu SETUP ALARMS ALARM DOOR 5 preset time in the user menu has elapsed SETUP ALARMS CLOCK ALARM the...

Page 24: ...of power shortage after which the program will be resumed decrease temp from 0 5 C to 25 C or off max drop below set temperature after which the program will continue increase temp from 0 5 C to 25 C...

Page 25: ...tures above 8 C the air automatically defrosts the ice cover defrosting is self operating 2 In temperatures below 8 C the evaporator may be covered in ice and the device should be defrosted manually I...

Page 26: ...s steel INOX we recommend using cleaning solution dedicated par ticularly to stainless steel material It preserves the steel surface from permanent stains and at the same time retains aesthetic appear...

Page 27: ...Open the door of the device and wait for the frost to melt take out the shelves and start cleaning the device 3 Only water or water with mild detergent should be used 4 Having finished cleaning you s...

Page 28: ...lean 4 Is the device placed in direct sunlight 5 Is there any heat source near the device 6 Are there too many objects inside the device that are not cool enough Water is condensing inside the chamber...

Page 29: ...the top Insert efficient T6 3 230V fuse push the fuse drawer into place Connect the power cable to the unit and then to a power outlet Turn on the device 17 PRODUCT IDENTIFICATION LABEL The product i...

Page 30: ...emedies IT IS EXPRESSLY AGREED THAT THIS WARRANTY WILL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE WARRANTY OF MERCHANTA BILITY Compliance with local laws and regulations The customer i...

Page 31: ...interior stainless steel housing powder coated sheet overal dims mm width 570 620 620 height 600 860 1260 depth 680 650 650 internal dims 2 mm width 470 520 520 height 430 660 1060 depth 300 410 410 w...

Page 32: ...Instruction Manual INCU Line Cooled Page 32 32...