19
From the UTILITY MODE select ‘SAMPLING SITE’ function.
SAMPLING SITE
MM/DD/YY - HH:MM
Press ‘ENTER’ to modify the sampling site
SAMPLING SITE
XXXX
Select the desired character using the “
▲
” or “
▼
” arrows and confirm with ‘ENTER’.
Select the next desired characters and confirm them by pressing ‘ENTER’.
The selected values are stored in the memory and it will be kept also after the instrument has been switched off.
‘CALIBRATION DUE’ (optional)
The SAS air sampler can be equipped with a calibration reminder feature. With this function the user can see when
the next calibration should be performed.
From the UTILITY MODE select the ‘CALIBRATION DUE’.
CALIBRATION DUE
MM/DD/YY - HH:MM
Press ‘ENTER’ to see the next calibration date:
NEXT CAL
MM/YYYY
The display shows the month in which the calibration should be performed.
One month before the calibration is due; the SAS will display the following warning message:
CALIBRATION DUE
MM/YYYY
When this message is displayed, press ‘ENTER’ to acknowledge.
When the calibration has expired, the SAS will display the following warning message:
CAL EXPIRED
When this message is displayed, press ‘ENTER’ to acknowledge.
The calibration timer can be reset only by a qualified technician.
List of system messages
Besides the messages described in the previous paragraph, the following messages can be displayed:
**LOW BATTERY**
Battery must be recharged
MOTOR ERROR
Motor not working or not connected
ENGLISH