CONTROL PANEL OVERVIEW
4.1
Power Switch:
The main power I/O (on/off) switch controls all power to the unit
and must be in the I/ON position before any systems are operational.
4.2 Main Temperature Control:
This control consists of the digital display and
Up/Down arrow pads for inputting set point temperatures and calibration.
4.3 Overtemperature
Thermostat:
This controller is equipped with an adjustment
knob and a graduated dial marked 1 to 10. Completely independent of the Main
Temperature Controller, the Overtemperature Thermostat guards against any
failure of the Main Temperature Controller which would allow temperature to rise
past set point. If temperature rises to the Overtemperature set point, the
Thermostat takes control of the heating element and allows continued use of the
incubator until the problem can be resolved or service can be arranged.
4.4 HEATING
Light:
This light is on when the unit is heating up to set point and is
blinking when controlling temperature at set point.
4.5 OVERTEMP Light:
This light comes on when the High Limit Thermostat is
activated. Under normal operating conditions this light should never come on.
4.6 Circuit
Breaker:
Located adjacent to the power cord, the breaker, when tripped,
is manually resettable by pushing in the extended button and offers protection
against power source variations.
4.7
Water Low Light:
When lit, this indicator shows that the water level in the jacket
is low, and DISTILLED WATER should be added until the indicator turns off.