36
Resetting your 9150
If the above troubleshooting tips do not correct the problem the following might help.
Restart the Meter by pressing the on/off key (2 sec).
•
Switch connections between a faulty input and the nearby input a see if the problem follows the load cell
or the input. Do forget to switch back the connections when the test is done.
•
Replace the Smart box. Use the restore function (Maintenance) to give the new Smart box the same
configuration and calibration as the old one.