33
8. Error codes
Note that for many of the texts below the Smart box number or channel / axle will be added.
Error #
Error text
Explanation
3
Instrument not in normal state!
(Err.3)
This is a fatal error with your Smart box. Contact your
dealer.
5
Input signal over range! (Err.5)
The mV/V input signal from the load cell(s) to the Smart box
is above approximately 4 mV/V. Check wiring, load cell.
7
Input signal under range! (Err.7)
The mV/V input signal from the load cell(s) is below
approximately -4 mV/V. Check wiring, load cell.
10
Low excitation voltage! (Err.10)
The load cell excitation voltage for the specified input is
below approx. 3.8V. Check load cell and wiring.
16
No valid weight data! (Err.16)
Indicates a Meter to Smart box communication error. Check
wiring between Meter and Smart boxes. Especially cables
between truck and trailer with connectors should be
checked.
17
Channel weight calc. error! (Err.17)
A channel/axle load calculation error occurred. Most likely
due to errors in true axle load parameter values entered at
configuration. Either typing error or a measuring error.
18
Channel weight out of range! (Err.18) Channel weight is exceeding 6 digits (incl. sign). Probably
due to calibration errors. Errors during tare or full load entry.
Summary of Contents for SI Onboard 9150
Page 2: ...p n 6190210 Rev B 5 5 21 JG...
Page 5: ...3 Use this flowchart as a basic guide and reference for setting up your 9150 system...
Page 36: ......
Page 41: ......
Page 45: ...39 10 Technical Data...